Literature DB >> 1745124

Spoiling of transverse magnetization in steady-state sequences.

Y Zur1, M L Wood, L J Neuringer.   

Abstract

A detailed analysis is presented of a method to eliminate transverse magnetization prior to each rf excitation in pulse sequences with TR less than T2. It is shown that artifact-free images with high T1 contrast can be obtained only if a phase shift that is incremented during each TR interval is applied to the transverse magnetization. Computer simulations are used to show that when this phase increment is 117 degrees, the steady-state transverse magnetization prior to each rf pulse is nulled over a wide range of T1, T2, and rf tip angles, resulting in optimal T1 contrast. Such nulling of steady-state transverse magnetization cannot be obtained by using large gradient pulses, or gradients of random or linearly incremented amplitude. Images of phantoms and human subjects confirm the theoretical predictions.

Entities:  

Mesh:

Year:  1991        PMID: 1745124     DOI: 10.1002/mrm.1910210210

Source DB:  PubMed          Journal:  Magn Reson Med        ISSN: 0740-3194            Impact factor:   4.668


  67 in total

1.  Physiological noise effects on the flip angle selection in BOLD fMRI.

Authors:  J Gonzalez-Castillo; V Roopchansingh; P A Bandettini; J Bodurka
Journal:  Neuroimage       Date:  2010-11-10       Impact factor: 6.556

2.  Assessment of Blood Hemodynamics by USPIO-Induced R(1) Changes in MRI of Murine Colon Carcinoma.

Authors:  Giulio Gambarota; H W M van Laarhoven; M Philippens; W J M Peeters; P Rijken; A van der Kogel; C J A Punt; A Heerschap
Journal:  Appl Magn Reson       Date:  2010-03-28       Impact factor: 0.831

3.  S5FP: spectrally selective suppression with steady state free precession.

Authors:  J A Derbyshire; D A Herzka; E R McVeigh
Journal:  Magn Reson Med       Date:  2005-10       Impact factor: 4.668

4.  Mapping the MRI voxel volume in which thermal noise matches physiological noise--implications for fMRI.

Authors:  J Bodurka; F Ye; N Petridou; K Murphy; P A Bandettini
Journal:  Neuroimage       Date:  2006-11-13       Impact factor: 6.556

5.  Improved half RF slice selectivity in the presence of eddy currents with out-of-slice saturation.

Authors:  Sonal Josan; Elena Kaye; John M Pauly; Bruce L Daniel; Kim Butts Pauly
Journal:  Magn Reson Med       Date:  2009-05       Impact factor: 4.668

6.  More than BOLD: Dual-spin populations create functional contrast.

Authors:  Amanda J Taylor; Jung H Kim; Vimal Singh; Elizabeth J Halfen; Josef Pfeuffer; David Ress
Journal:  Magn Reson Med       Date:  2019-08-18       Impact factor: 4.668

7.  Contrast-modified gradient echo imaging using rotary echo preparatory pulses.

Authors:  X P Zhu; P B Chilvers; C E Hutchinson; G A Morris; J M Hawnaur; J E Adams; C J Taylor
Journal:  MAGMA       Date:  1997-09       Impact factor: 2.310

8.  High-grade soft-tissue sarcoma: optimizing injection improves MRI evaluation of tumor response.

Authors:  Amandine Crombé; François Le Loarer; François Cornelis; Eberhardt Stoeckle; Xavier Buy; Sophie Cousin; Antoine Italiano; Michèle Kind
Journal:  Eur Radiol       Date:  2018-07-23       Impact factor: 5.315

9.  Optimization of magnetization-prepared rapid gradient echo (MP-RAGE) sequence for neonatal brain MRI.

Authors:  Lili He; Jinghua Wang; Zhong-Lin Lu; Beth M Kline-Fath; Nehal A Parikh
Journal:  Pediatr Radiol       Date:  2018-05-02

10.  Improved spoiling efficiency in dynamic RF-spoiled imaging by ghost phase modulation and temporal filtering.

Authors:  Jon-Fredrik Nielsen; Douglas C Noll
Journal:  Magn Reson Med       Date:  2015-07-07       Impact factor: 4.668

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.