Literature DB >> 17441671

Early removal of a Maverick disc prosthesis: surgical findings and morphological changes.

Jens François1, Ronald Coessens, Philippe Lauweryns.   

Abstract

We report the case of a patient who underwent explantation of a Maverick total disc prosthesis at the L5S1 level because of severe persisting pain one year after initial implantation. Dynamic radiographic imaging studies showed good position and size of the prosthesis and no evidence of loosening. Intraoperatively residual mobility at the proximal bone-prosthesis interface was detected, as well as gross metallosis around the articulation of the total disc prosthesis. A safe and straightforward technique for the extraction of a Maverick prosthesis using a hooked instrument to hook-on the keel of the device is reported. After removal of the device, an anterior lumbar interbody fusion with subsequent posterior pedicle screw fixation and posterior bone grafting with autologous iliac bone was performed (360 degrees fusion). During this posterior procedure, severe L5S1 facet joint arthrosis was observed. Early clinical and radiographic results were good with excellent patient satisfaction.

Entities:  

Mesh:

Year:  2007        PMID: 17441671

Source DB:  PubMed          Journal:  Acta Orthop Belg        ISSN: 0001-6462            Impact factor:   0.500


  12 in total

Review 1.  Design concepts in lumbar total disc arthroplasty.

Authors:  Fabio Galbusera; Chiara M Bellini; Thomas Zweig; Stephen Ferguson; Manuela T Raimondi; Claudio Lamartina; Marco Brayda-Bruno; Maurizio Fornari
Journal:  Eur Spine J       Date:  2008-10-23       Impact factor: 3.134

2.  Assessment of epidural versus intradiscal biocompatibility of PEEK implant debris: an in vivo rabbit model.

Authors:  Nadim J Hallab; Qi-Bin Bao; Tim Brown
Journal:  Eur Spine J       Date:  2013-08-31       Impact factor: 3.134

Review 3.  Which design and biomaterial factors affect clinical wear performance of total disc replacements? A systematic review.

Authors:  Sai Y Veruva; Marla J Steinbeck; Jeffrey Toth; Dominik D Alexander; Steven M Kurtz
Journal:  Clin Orthop Relat Res       Date:  2014-12       Impact factor: 4.176

4.  Corrosion of Harrington rod in idiopathic scoliosis: long-term effects.

Authors:  Beth Sherman; Tanya Crowell
Journal:  Eur Spine J       Date:  2017-06-17       Impact factor: 3.134

Review 5.  Metallosis after traumatic loosening of Bryan cervical disc arthroplasty: a case report and literature review.

Authors:  Chih-Chan Yang; Chien-Lun Tang; Chung-Yuh Tzeng; Hsi-Kai Tsou
Journal:  Eur Spine J       Date:  2017-11-21       Impact factor: 3.134

Review 6.  Proper animal experimental designs for preclinical research of biomaterials for intervertebral disc regeneration.

Authors:  Yizhong Peng; Xiangcheng Qing; Hongyang Shu; Shuo Tian; Wenbo Yang; Songfeng Chen; Hui Lin; Xiao Lv; Lei Zhao; Xi Chen; Feifei Pu; Donghua Huang; Xu Cao; Zengwu Shao
Journal:  Biomater Transl       Date:  2021-06-28

Review 7.  Spinal metallosis: a systematic review.

Authors:  Yoni Goldenberg; Jin W Tee; Cesar M Salinas-La Rosa; Michael Murphy
Journal:  Eur Spine J       Date:  2016-01-05       Impact factor: 3.134

8.  Prospective study on serum metal levels in patients with metal-on-metal lumbar disc arthroplasty.

Authors:  Matthew F Gornet; J K Burkus; M L Harper; F W Chan; A K Skipor; J J Jacobs
Journal:  Eur Spine J       Date:  2012-11-20       Impact factor: 3.134

9.  Interaction of micron and nano-sized particles with cells of the dura mater.

Authors:  Iraklis Papageorgiou; Rainy Marsh; Joanne L Tipper; Richard M Hall; John Fisher; Eileen Ingham
Journal:  J Biomed Mater Res B Appl Biomater       Date:  2014-03-06       Impact factor: 3.368

10.  Spine imaging after lumbar disc replacement: pitfalls and current recommendations.

Authors:  Yohan Robinson; Bengt Sandén
Journal:  Patient Saf Surg       Date:  2009-07-20
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.