PURPOSE: To evaluate the ability of a custom-built coil design to provide improved signal-to-noise ratio (SNR) and less signal drop with increasing depth at the carotid artery. MATERIALS AND METHODS: Phased-array surface coils can provide a high SNR to image the carotid vessel wall. However, given the required field-of-view (FOV) and penetration depth, these coils show either a fast signal drop with increasing depth or a moderate SNR at increased coil size. A quadrature surface coil (a butterfly coil in conjunction with a linear single-loop coil) was compared with a phased-array coil in phantom and human studies. RESULTS: The phantom studies showed that the quadrature coil has better SNR over the required FOV than a standard phased-array coil (26% at 3 cm depth). CONCLUSION: The quadrature coil enables better image quality to be achieved. (c) 2007 Wiley-Liss, Inc.
PURPOSE: To evaluate the ability of a custom-built coil design to provide improved signal-to-noise ratio (SNR) and less signal drop with increasing depth at the carotid artery. MATERIALS AND METHODS: Phased-array surface coils can provide a high SNR to image the carotid vessel wall. However, given the required field-of-view (FOV) and penetration depth, these coils show either a fast signal drop with increasing depth or a moderate SNR at increased coil size. A quadrature surface coil (a butterfly coil in conjunction with a linear single-loop coil) was compared with a phased-array coil in phantom and human studies. RESULTS: The phantom studies showed that the quadrature coil has better SNR over the required FOV than a standard phased-array coil (26% at 3 cm depth). CONCLUSION: The quadrature coil enables better image quality to be achieved. (c) 2007 Wiley-Liss, Inc.
Authors: Qinwei Zhang; Bram F Coolen; Sandra van den Berg; Gyula Kotek; Debra S Rivera; Dennis W J Klomp; Gustav J Strijkers; Aart J Nederveen Journal: PLoS One Date: 2019-03-04 Impact factor: 3.240