Cameron J Pedersen1, David Huang, Mark A Shure, Andrew M Rollins.
Abstract
Single-beam laser Doppler measurements of flow velocity are only sensitive to the velocity component parallel to the optical axis. We describe a simple modification to a standard Doppler optical coherence tomography (OCT) system using a single sample beam that provides velocity information from multiple angles within the beam. By introducing a glass plate midway into the OCT beam path, the sample beam is divided into several components, each with a different group delay and each providing a separate interferogram with its own effective Doppler angle. By combining the Doppler shift measured in each of these component interferograms, the flow velocity vector is fully determined.Mesh:
Year: 2007 PMID: 17392903 DOI: 10.1364/ol.32.000506
Source DB: PubMed Journal: Opt Lett ISSN: 0146-9592 Impact factor: 3.776