Literature DB >> 17364656

Seed localization and TRUS-fluoroscopy fusion for intraoperative prostate brachytherapy dosimetry.

Yi Su1, Brian J Davis, Keith M Furutani, Michael G Herman, Richard A Robb.   

Abstract

OBJECTIVE: To develop and evaluate an integrated approach to intra-operative dosimetry for permanent prostate brachytherapy (PPB) by combining a fluoroscopy-based seed localization routine with a transrectal ultrasound (TRUS)-to-fluoroscopy fusion technique.
MATERIALS AND METHODS: Three-dimensional seed coordinates are reconstructed based on the two-dimensional seed locations identified from three fluoroscopic images acquired at different angles. A seed-based registration approach was examined in both simulation and phantom studies to register the seed locations identified from the fluoroscopic images to the TRUS images. Dose parameters were then evaluated and compared to CT-based dosimetry from a patient dataset.
RESULTS: Less than 0.2% error in the D90 value was observed using the TRUS-fluoroscopy image-fusion-based method relative to the CT-based post-implantation dosimetry. In the phantom study, an average distance of 3 mm was observed between the seeds identified from TRUS and the reconstructed seeds at registration. Isodose contours were displayed superimposed on the TRUS images.
CONCLUSIONS: Promising results were observed in this preliminary study of a TRUS-fluoroscopy fusion-based brachytherapy dosimetry analysis method, suggesting that the method is highly sensitive and calculates clinically relevant dosimetry, including the prostate D90. Further validation of the method is required for eventual clinical application.

Entities:  

Mesh:

Year:  2007        PMID: 17364656     DOI: 10.3109/10929080601168239

Source DB:  PubMed          Journal:  Comput Aided Surg        ISSN: 1092-9088


  7 in total

1.  INTENSITY-BASED REGISTRATION OF PROSTATE BRACHYTHERAPY IMPLANTS AND ULTRASOUND.

Authors:  Z Karimaghaloo; G Fichtinger; D G Gobbi; E C Burdette; R N Rohling; P Abolmaesumi
Journal:  Proc IEEE Int Symp Biomed Imaging       Date:  2008-06-13

2.  Feasibility of vibro-acoustography with a quasi-2D ultrasound array transducer for detection and localizing of permanent prostate brachytherapy seeds: a pilot ex vivo study.

Authors:  Mohammad Mehrmohammadi; Azra Alizad; Randall R Kinnick; Brian J Davis; Mostafa Fatemi
Journal:  Med Phys       Date:  2014-09       Impact factor: 4.071

3.  Intra-operative Localization of Brachytherapy Implants Using Intensity-based Registration.

Authors:  Z Karimaghaloo; P Abolmaesumi; N Ahmidi; T K Chen; D G Gobbi; G Fichtinger
Journal:  Proc SPIE Int Soc Opt Eng       Date:  2009-01-01

4.  Ultrasound-fluoroscopy registration for prostate brachytherapy dosimetry.

Authors:  Ehsan Dehghan; Junghoon Lee; Pascal Fallavollita; Nathanael Kuo; Anton Deguet; Yi Le; E Clif Burdette; Danny Y Song; Jerry L Prince; Gabor Fichtinger
Journal:  Med Image Anal       Date:  2012-06-16       Impact factor: 8.545

5.  Point-to-volume registration of prostate implants to ultrasound.

Authors:  Ehsan Dehghan; Junghoon Lee; Pascal Fallavollita; Nathanael Kuo; Anton Deguet; E Clif Burdette; Danny Song; Jerry L Prince; Gabor Fichtinger
Journal:  Med Image Comput Comput Assist Interv       Date:  2011

6.  Quantification of edematic effects in prostate brachytherapy interventions.

Authors:  Mohamed Hefny; Purang Abolmaesumi; Zahra Karimaghaloo; David G Gobbi; Randy Ellis; Gabor Fichtinger
Journal:  Med Image Comput Comput Assist Interv       Date:  2008

Review 7.  Image fusion techniques in permanent seed implantation.

Authors:  Alfredo Polo
Journal:  J Contemp Brachytherapy       Date:  2010-10-13
  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.