| Literature DB >> 1731664 |
H M Spotnitz1, G Y Ott, J T Bigger, J S Steinberg, F Livelli.
Abstract
Experience with combined transvenous pacemaker and implantable cardioverter-defibrillator insertion in 21 patients is described. Special techniques are needed to avoid potentially lethal pacemaker-implantable cardioverter-defibrillator interaction. Separation between leads for the two devices should be maximized. The electrophysiologic criteria for successful device function can be met, even when some leads for both devices must be placed by a transvenous approach.Entities:
Mesh:
Year: 1992 PMID: 1731664 DOI: 10.1016/0003-4975(92)91327-6
Source DB: PubMed Journal: Ann Thorac Surg ISSN: 0003-4975 Impact factor: 4.330