| Literature DB >> 1730983 |
S H Manglos1, D A Bassano, F D Thomas, Z D Grossman.
Abstract
Radionuclide transmission CT generated on a rotating gamma camera can improve SPECT imaging by providing attenuation maps for attenuation compensation and for anatomical correlation. This paper demonstrates the feasibility and high quality of cone-beam transmission CT (CB-CT) of human subjects, in comparison to conventional parallel-ray CT, and evaluates some possible imaging protocols. Two CB-CT implementation modes, with a cone-beam collimator and without any collimator, were evaluated. Three human subjects of different dimensions were imaged. For the two smaller subjects, the CB-CT images were dramatically superior, in terms of noise and resolution, to those obtained with a parallel-ray geometry. The image noise was less by a factor of 6. CB-CT linear attenuation coefficients were found to be in close agreement with published values for various tissues. For the largest subject, image truncation produced a ring artifact at the edge, but inside the artifact, the image quality was still very good. Cone-beam images obtained without any collimator were acceptable, but photon scatter degraded the image contrast.Entities:
Mesh:
Substances:
Year: 1992 PMID: 1730983
Source DB: PubMed Journal: J Nucl Med ISSN: 0161-5505 Impact factor: 10.057