Anatomic factors associated with acute endograft collapse after Gore TAG treatment of thoracic aortic dissection or traumatic rupture.

Bart E Muhs1, Ron Balm, Geoffrey H White, Hence J M Verhagen.   

Abstract

OBJECTIVE: The potentially devastating complication of total or near total thoracic endoprosthesis collapse has been described with the TAG device (W. L. Gore & Associates, Flagstaff, Ariz). This rare complication has resulted in a warning to clinicians and speculation about the etiology of this problem. This report evaluates potential causative anatomic factors that may increase the probability of endoprosthesis collapse in patients undergoing endovascular thoracic aneurysm repair (TEVAR).
METHODS: Preoperative and postoperative computed tomography scans were collected worldwide representing six patients who had experienced radiologically confirmed TAG endoprosthesis collapse. These were compared with a matched cohort of five patients with a TAG endoprosthesis in the same anatomic position in which no collapse occurred. Anatomic variables of aortic arch angulation, apposition, intraluminal lip length, proximal aortic diameter, distal aortic diameter, intragraft aortic diameter, percentage of oversizing, and angle of the proximal endograft to the aortic arch were compared between groups. Differences between groups were determined using the Student t test, with P < .05 considered significant.
RESULTS: The two groups (collapse vs no collapse) were evenly matched demographically, and all underwent endoluminal treatment with the TAG device, with no differences in gender, graft position in the aorta, operative indication, or age (P = NS). Distal sealing zone aortic diameter +/- standard deviation of 18.9 +/- 1.7 mm vs 22.7 +/- 2.7 mm and minimum aortic diameter within the endograft of 18.6 +/- 1.7 mm vs 22.4 +/- 3.1 mm predicted collapse (P < .05). Proximal aortic diameter, apposition, intraluminal lip length, aortic arch angle, and angle of proximal endograft to aortic arch did not predict collapse (P = NS).
CONCLUSION: Thoracic endograft collapse is an exceedingly rare event. In this series, endoprosthesis collapse occurred in patients who were treated outside the manufacturer's instructions for use for minimum required aortic diameter. Although distal aortic diameter and minimum intragraft aortic diameter predicted collapse, other variables may also influence this complication but were not significant owing to potential type II statistical errors. In the future, caution should be exercised when contemplating TEVAR in patients with small (<23 mm) aortic diameters.

Entities:  

Mesh:

Year:  2007        PMID: 17306949     DOI: 10.1016/j.jvs.2006.12.023

Source DB:  PubMed          Journal:  J Vasc Surg        ISSN: 0741-5214            Impact factor:   4.268


  12 in total

1.  Unusual complications of endovascular repair of the thoracic aorta: MDCT findings.

Authors:  T Valente; G Rossi; F Lassandro; G Rea; M Marino; G Dialetto; R Muto; M Scaglione
Journal:  Radiol Med       Date:  2012-01-07       Impact factor: 3.469

2.  Endovascular repair of a traumatic thoracic aortic injury resulting in late aortic dilatation.

Authors:  Kumi Ozaki; Osamu Matsui; Hiroshi Ohtake; Go Watanabe; Junichiro Anada
Journal:  Int J Angiol       Date:  2013-06

3.  Effects of longitudinal pre-stretch on the mechanics of human aorta before and after thoracic endovascular aortic repair (TEVAR) in trauma patients.

Authors:  Anastasia Desyatova; Jason MacTaggart; Alexey Kamenskiy
Journal:  Biomech Model Mechanobiol       Date:  2019-09-05

4.  Endovascular repair of thoracic aortic injury: current thoughts and technical considerations.

Authors:  W Darrin Clouse
Journal:  Semin Intervent Radiol       Date:  2010-03       Impact factor: 1.513

5.  Current management of traumatic rupture of the descending thoracic aorta.

Authors:  Riyad Karmy-Jones; Nichole Jackson; William Long; Alan Simeone
Journal:  Curr Cardiol Rev       Date:  2009-08

6.  Endovascular and open repair for blunt aortic injury, treated in one clinical institution in Brazil: a case series.

Authors:  Igor Rafael Sincos; Ricardo Aun; Sergio Quilici Belczak; Luciano Dias Nascimento; Boulanger Mioto Netto; Ivan Casella; Erasmo Simao da Silva; Pedro Puech-Leão
Journal:  Clinics (Sao Paulo)       Date:  2011       Impact factor: 2.365

7.  Delayed endovascular treatment of descending aorta stent graft collapse in a patient treated for post- traumatic aortic rupture: a case report.

Authors:  Giovanni Nano; Daniela Mazzaccaro; Giovanni Malacrida; Maria Teresa Occhiuto; Silvia Stegher; Domenico G Tealdi
Journal:  J Cardiothorac Surg       Date:  2011-05-24       Impact factor: 1.637

8.  Traumatic aortic injury: does the anatomy of the aortic arch influence aortic trauma severity?

Authors:  Jacek Wojciechowski; Lukasz Znaniecki; Kamil Bury; Kamil Chwojnicki; Jan Rogowski
Journal:  Surg Today       Date:  2016-11-17       Impact factor: 2.549

9.  Aortic Isthmus Narrowing after Endovascular Repair of Acute Traumatic Aortic Transection.

Authors:  Joung Taek Kim; Young Sam Kim; Yong Han Yoon; Cheol Wong Kang; Wan Ki Baek; Do Hyun Kim
Journal:  Vasc Specialist Int       Date:  2018-09-30

10.  Infolding of fenestrated endovascular stent graft.

Authors:  Jason G E Zelt; Prasad Jetty; Adnan Hadziomerovic; Sudhir Nagpal
Journal:  J Vasc Surg Cases Innov Tech       Date:  2017-07-20
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.