[Quantification of coronary artery stenoses: comparison of 64-slice and dual source CT angiography with cardiac catheterization].

Stephanie Busch1, K Nikolaou, T Johnson, C Rist, A Knez, M Reiser, C Becker.   

Abstract

BACKGROUND: Until now stenoses of the coronary arteries have been evaluated visually with CT angiography. Therefore, the results were highly dependent on subjective factors inherent in the examiner. New software tools for semiquantitative analysis (CT-QCA, quantitative coronary assessment) might be adequate to improve the diagnostic accuracy und reproducibility.
MATERIAL AND METHODS: CTAs of 20 patients were analyzed. Ten patients each were evaluated using 64-slice CT (64SCT) and dual source CT (DSCT) (Somatom Sensation 64 and Somatom Definition, Siemens Medical Solutions, Forchheim), respectively. Two radiologists independently evaluated the data visually and with the help of a software tool (Syngo Circulation, Siemens Medical Solutions, Forchheim). The results of the quantitative assessment of the invasive heart catheterization served as the reference standard. Sensitivity and specificity as well as the correlation coefficient, the systematic error, and the interobserver agreement (kappa) were determined.
RESULTS: In each of both patient groups 12 stenoses were detected. For the detection of stenoses >75%, sensitivity and specificity for the visual evaluation using the 64SCT were 100% and 90%, and with the CT-QCA both were 100%. For the DSCT sensitivity and specificity were 100% for both the visual and semiautomated evaluation. The Bland-Altman plot of the results of the 64SCT showed an overestimation of 3.3% (+/-62.7%/56.2%) compared to the heart catheterization. The results of the DSCT exhibited an overestimation of 6.2% (+/-33.1%/19.8%). The interobserver agreement of the CT-QCA and the visual evaluation showed a kappa value of 0.75 and for DSCT of 1.0.
CONCLUSION: The results showed a good correlation of grading stenosis between the software-assisted evaluation and the results of the coronary catheter angiography. The promising results of the DSCT are due to a superior temporal resolution compared to the 64SCT. Confirmation of these data by trials in larger patient collectives is warranted.

Entities:  

Mesh:

Year:  2007        PMID: 17287946     DOI: 10.1007/s00117-007-1476-x

Source DB:  PubMed          Journal:  Radiologe        ISSN: 0033-832X            Impact factor:   0.635


  13 in total

1.  Advances in cardiac CT imaging: 64-slice scanner.

Authors:  Konstantin Nikolaou; Thomas Flohr; Andreas Knez; Carsten Rist; Bernd Wintersperger; Thorsten Johnson; Maximilian F Reiser; Christoph R Becker
Journal:  Int J Cardiovasc Imaging       Date:  2004-12       Impact factor: 2.357

2.  Comparison of the degree of coronary stenoses by multidetector computed tomography versus by quantitative coronary angiography.

Authors:  Ricardo C Cury; Eugene V Pomerantsev; Maros Ferencik; Udo Hoffmann; Koen Nieman; Fabian Moselewski; Suhny Abbara; Ik-Kyung Jang; Thomas J Brady; Stephan Achenbach
Journal:  Am J Cardiol       Date:  2005-09-15       Impact factor: 2.778

3.  The limitations of coronary angiography: identification of a critical coronary stenosis using intravascular ultrasound.

Authors:  R A Swallow; I A Court; A L Calver; N P Curzen
Journal:  Int J Cardiol       Date:  2006-01-04       Impact factor: 4.164

4.  Dual-source CT cardiac imaging: initial experience.

Authors:  Thorsten R C Johnson; Konstantin Nikolaou; Bernd J Wintersperger; Alexander W Leber; Franz von Ziegler; Carsten Rist; Sonja Buhmann; Andreas Knez; Maximilian F Reiser; Christoph R Becker
Journal:  Eur Radiol       Date:  2006-05-13       Impact factor: 5.315

5.  Accuracy of 64-MDCT in the diagnosis of ischemic heart disease.

Authors:  Konstantin Nikolaou; Andreas Knez; Carsten Rist; Bernd J Wintersperger; Alexander Leber; Thorsten Johnson; Maximilian F Reiser; Christoph R Becker
Journal:  AJR Am J Roentgenol       Date:  2006-07       Impact factor: 3.959

6.  Quantitative parameters of image quality in 64-slice computed tomography angiography of the coronary arteries.

Authors:  Maros Ferencik; Cesar H Nomura; Pal Maurovich-Horvat; Udo Hoffmann; Antonio J Pena; Ricardo C Cury; Suhny Abbara; Koen Nieman; Umaima Fatima; Stephan Achenbach; Thomas J Brady
Journal:  Eur J Radiol       Date:  2006-01-24       Impact factor: 3.528

7.  Multislice spiral computed tomography coronary angiography in patients with stable angina pectoris.

Authors:  Nico R Mollet; Filippo Cademartiri; Koen Nieman; Francesco Saia; Pedro A Lemos; Eugene P McFadden; Peter M T Pattynama; Patrick W Serruys; Gabriel P Krestin; Pim J de Feyter
Journal:  J Am Coll Cardiol       Date:  2004-06-16       Impact factor: 24.094

8.  Accuracy of thin-slice computed tomography in the detection of coronary stenoses.

Authors:  Eugenio Martuscelli; Andrea Romagnoli; Alessia D'Eliseo; Cinzia Razzini; Marco Tomassini; Massimiliano Sperandio; Giovanni Simonetti; Francesco Romeo
Journal:  Eur Heart J       Date:  2004-06       Impact factor: 29.983

9.  Significance of automated stenosis detection during quantitative angiography. Insights gained from intracoronary ultrasound imaging.

Authors:  J Escaned; J Baptista; C Di Mario; J Haase; Y Ozaki; D T Linker; P J de Feyter; J R Roelandt; P W Serruys
Journal:  Circulation       Date:  1996-09-01       Impact factor: 29.690

10.  Value of 3D-volume rendering in the assessment of coronary arteries with retrospectively ECG-gated multislice spiral CT.

Authors:  A H Mahnken; J E Wildberger; A M Sinha; K Dedden; S Stanzel; R Hoffmann; T Schmitz-Rode; R W Günther
Journal:  Acta Radiol       Date:  2003-05       Impact factor: 1.701

View more
  4 in total

1.  Multislice CT angiography in coronary artery disease: Technical developments, radiation dose and diagnostic value.

Authors:  Zhonghua Sun
Journal:  World J Cardiol       Date:  2010-10-26

2.  Dual source computed tomography: automated, visual or dual analysis?

Authors:  E E van der Wall; J H C Reiber
Journal:  Int J Cardiovasc Imaging       Date:  2008-11-27       Impact factor: 2.357

3.  The effect of iterative reconstruction on quantitative computed tomography assessment of coronary plaque composition.

Authors:  Richard A P Takx; Martin J Willemink; Hendrik M Nathoe; Arnold M R Schilham; Ricardo P J Budde; Pim A de Jong; Tim Leiner
Journal:  Int J Cardiovasc Imaging       Date:  2013-09-18       Impact factor: 2.357

4.  [Coronary heart-CT].

Authors:  N Faucheron; P Reimer
Journal:  Radiologe       Date:  2008-07       Impact factor: 0.635

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.