Literature DB >> 17275643

Polyethylene wear against alumina and zirconia heads in cemented total hip arthroplasty.

Bojian Liang1, Keiichi Kawanabe, Kentaro Ise, Hirokazu Iida, Takashi Nakamura.   

Abstract

We compared the polyethylene wear of acetabular sockets articulated with 22.225-mm alumina heads with the polyethylene wear of those articulated with 22.225-mm zirconia heads in cemented total hip arthroplasty during a mean follow-up period of 5.4 years. Using a computer-aided technique, we measured polyethylene wear radiologically in 46 hips with alumina heads and 58 hips with zirconia heads. The preoperative diagnosis in all cases was osteoarthritis. The mean linear wear rate and mean volumetric wear rate of polyethylene sockets against zirconia heads were 0.133 mm/y and 39.8 mm(3)/y, respectively, significantly greater (P < .01) than the wear rates against alumina heads (0.078 mm/y and 24.2 mm(3)/y, respectively). Age at operation, patient body weight as well as height, thickness of polyethylene, and socket abduction angle did not influence the wear rates. We speculate that the excessive polyethylene wear was caused by phase transformation of zirconia, leading to an increase of surface roughness.

Entities:  

Mesh:

Substances:

Year:  2007        PMID: 17275643     DOI: 10.1016/j.arth.2006.03.004

Source DB:  PubMed          Journal:  J Arthroplasty        ISSN: 0883-5403            Impact factor:   4.757


  8 in total

Review 1.  Manufacturers affect clinical results of THA with zirconia heads: a systematic review.

Authors:  Hiroyuki Yoshitomi; Satoru Shikata; Hiromu Ito; Takeo Nakayama; Takashi Nakamura
Journal:  Clin Orthop Relat Res       Date:  2009-01-31       Impact factor: 4.176

2.  In vitro biological and tribological properties of transparent magnesium aluminate (Spinel) and aluminum oxynitride (ALON®).

Authors:  Subhadip Bodhak; Vamsi Krishna Balla; Susmita Bose; Amit Bandyopadhyay; Uday Kashalikar; Santosh K Jha; Suri Sastri
Journal:  J Mater Sci Mater Med       Date:  2011-05-12       Impact factor: 3.896

Review 3.  Advances in zirconia toughened alumina biomaterials for total joint replacement.

Authors:  Steven M Kurtz; Sevi Kocagöz; Christina Arnholt; Roland Huet; Masaru Ueno; William L Walter
Journal:  J Mech Behav Biomed Mater       Date:  2013-04-10

4.  Similar wear in total hip arthroplasties with metallic or zirconia femoral heads.

Authors:  Randy M Cohn; Alejandro González Della Valle; Margaret Peterson; Charles N Cornell
Journal:  HSS J       Date:  2008-07-15

5.  Alumina-on-Polyethylene Bearing Surfaces in Total Hip Arthroplasty.

Authors:  Yup Lee Jung; Shin-Yoon Kim
Journal:  Open Orthop J       Date:  2010-02-11

6.  A bicentric approach evaluating the combination of a hemispheric cup with a novel ceramic head in total hip arthroplasty.

Authors:  André Busch; Dennis Wassenaar; Wolfgang Zinser; Marcus Jäger
Journal:  Orthop Rev (Pavia)       Date:  2021-03-31

7.  In vivo wear performance of highly cross-linked polyethylene vs. yttria stabilized zirconia and alumina stabilized zirconia at a mean seven-year follow-up.

Authors:  Masahiro Hasegawa; Akihiro Sudo
Journal:  BMC Musculoskelet Disord       Date:  2013-05-01       Impact factor: 2.362

8.  Catastrophic Failure of the Acetabular Polyethylene Liner in Ceramic-on-Polyethylene Total Hip Arthroplasty.

Authors:  Givenchy Manzano; Rayna A C Levin; Michael B Mayor; Ran Schwarzkopf
Journal:  J Orthop Case Rep       Date:  2014 Apr-Jun
  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.