| Literature DB >> 17271926 |
Haining Sun1, Shaokun Qiu, Shanshan Lou, Jinjun Liu, Changjun Li, Genmiao Jiang.
Abstract
This work presents a new method to reduce nonlinear artifacts in computed tomography (CT). Based on the traditional water-equivalent beam hardening correction method, a new systematic iterative algorithm has been designed to modify the original spectrum, which is under the influences of certain added filter materials and some unknown factors. By incorporating the characters of polychromatic beam hardening and the insensibility and nonuniformity of detectors into consideration, a new polynomial function curve is calculated. The curve can calibrate CT raw data and reduce the nonlinear artifacts, such as shading artifacts, dark artifacts, cupping artifacts and ring artifacts, in soft tissue. Comparing with the traditional water-equivalent correction, results show that this method can significantly improve the image quality. Meanwhile, the method is pre-processing and will not increase the normal reconstruction time. That is, all the time-consuming works can be done before scanning patients. However, it is still depending on the size of phantoms currently used, and more detailed works need to be done in the future.Entities:
Year: 2004 PMID: 17271926 DOI: 10.1109/IEMBS.2004.1403407
Source DB: PubMed Journal: Conf Proc IEEE Eng Med Biol Soc ISSN: 1557-170X