Reliability of confocal microscopy spectral imaging systems: use of multispectral beads.

Robert M Zucker1, Paul Rigby, Ian Clements, Wendy Salmon, Michael Chua.   

Abstract

BACKGROUND: There is a need for a standardized, impartial calibration, and validation protocol on confocal spectral imaging (CSI) microscope systems. To achieve this goal, it is necessary to have testing tools to provide a reproducible way to evaluate instrument performance.
METHODS: We evaluated fluorescent spectral beads (FocalCheck) from Molecular Probes/Invitrogen that consist of four pairs with emissions between 500 and 725 nm and a europium macrocycle quantum dye bead. These bead tools compliment our previously published protocol for testing spectral imaging systems that used an inexpensive multi-ion discharge lamp (MIDL) that contains Hg(+), Ar(+), and inorganic fluorophores that emits distinct, stable spectral features.
RESULTS: We acquired the spectra of the FocalCheck beads on a Zeiss 510 Meta, a Leica TCS-SP1, a Leica SP2 AOBS, an Olympus FV 1000, and a Nikon C1Si confocal systems and a PARISS microscopic spectral system and of the europium beads on the Leica TCS-SP1 and PARISS spectral imaging systems. A lack of performance with some equipment between 650 and 750 nm was identified using the far red pair of the FocalCheck beads. The position of the slider in front of PMT 2 that reflects light into PMT 1 and PMT 3 affected the measurement of all bead intensities. Unmixing algorithms were used to separate beads with different fluorochromes and separate two fluorochromes on the same bead.
CONCLUSIONS: The FocalCheck multi-spectral beads yielded similar profiles on four CSI systems and a PARISS spectral system. The utilization of the spectral FocalCheck beads is helpful to evaluate proper spectral performance, especially in the far red region. Europium beads provide a very narrow spectrum that can help to identify machines that have spectral problems. The dyes located on individual beads or mixed together in ring-core configuration can be used as test particles to demonstrate spectral unmixing with various algorithms.

Mesh:

Substances:

Year:  2007        PMID: 17266146     DOI: 10.1002/cyto.a.20371

Source DB:  PubMed          Journal:  Cytometry A        ISSN: 1552-4922            Impact factor:   4.355


  10 in total

1.  Thin-film tunable filters for hyperspectral fluorescence microscopy.

Authors:  Peter Favreau; Clarissa Hernandez; Ashley Stringfellow Lindsey; Diego F Alvarez; Thomas Rich; Prashant Prabhat; Silas J Leavesley
Journal:  J Biomed Opt       Date:  2014-01       Impact factor: 3.170

2.  Hierarchical functional gradients of pH-responsive self-assembled monolayers using dynamic covalent chemistry on surfaces.

Authors:  Lara Tauk; André P Schröder; Gero Decher; Nicolas Giuseppone
Journal:  Nat Chem       Date:  2009-10-23       Impact factor: 24.427

3.  Microscopy is better in color: development of a streamlined spectral light path for real-time multiplex fluorescence microscopy.

Authors:  Craig M Browning; Samantha Mayes; Samuel A Mayes; Thomas C Rich; Silas J Leavesley
Journal:  Biomed Opt Express       Date:  2022-06-07       Impact factor: 3.562

4.  Quality assessment in light microscopy for routine use through simple tools and robust metrics.

Authors:  Orestis Faklaris; Leslie Bancel-Vallée; Aurélien Dauphin; Baptiste Monterroso; Perrine Frère; David Geny; Tudor Manoliu; Sylvain de Rossi; Fabrice P Cordelières; Damien Schapman; Roland Nitschke; Julien Cau; Thomas Guilbert
Journal:  J Cell Biol       Date:  2022-09-29       Impact factor: 8.077

5.  An integrated imaging approach to the study of oxidative stress generation by mitochondrial dysfunction in living cells.

Authors:  Wan-Yun Cheng; Haiyan Tong; Evan W Miller; Christopher J Chang; James Remington; Robert M Zucker; Philip A Bromberg; James M Samet; Thomas P J Hofer
Journal:  Environ Health Perspect       Date:  2010-04-22       Impact factor: 9.031

6.  ConfocalCheck--a software tool for the automated monitoring of confocal microscope performance.

Authors:  Keng Imm Hng; Dirk Dormann
Journal:  PLoS One       Date:  2013-11-05       Impact factor: 3.240

7.  Darkfield-confocal microscopy detection of nanoscale particle internalization by human lung cells.

Authors:  Eugene A Gibbs-Flournoy; Philip A Bromberg; Thomas P J Hofer; James M Samet; Robert M Zucker
Journal:  Part Fibre Toxicol       Date:  2011-01-19       Impact factor: 9.400

8.  Fluorescence microspectroscopy as a tool to study mechanism of nanoparticles delivery into living cancer cells.

Authors:  Zoran Arsov; Iztok Urbančič; Maja Garvas; Daniele Biglino; Ajasja Ljubetič; Tilen Koklič; Janez Strancar
Journal:  Biomed Opt Express       Date:  2011-06-29       Impact factor: 3.732

9.  Spectral unmixing: analysis of performance in the olfactory bulb in vivo.

Authors:  Mathieu Ducros; Laurent Moreaux; Jonathan Bradley; Pascale Tiret; Oliver Griesbeck; Serge Charpak
Journal:  PLoS One       Date:  2009-02-09       Impact factor: 3.240

10.  eSIP: A Novel Solution-Based Sectioned Image Property Approach for Microscope Calibration.

Authors:  Malte Butzlaff; Arwed Weigel; Evgeni Ponimaskin; Andre Zeug
Journal:  PLoS One       Date:  2015-08-05       Impact factor: 3.240

  10 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.