Torsional stability of intramedullary compression nails: tibial osteotomy model.

Nicholas A T Brown1, Nathaniel A Bryan, Peter M Stevens.   

Abstract

BACKGROUND: Studies comparing intramedullary compression nailing to conventional dynamized intramedullary nailing contend that better clinical outcomes of intramedullary compression techniques result from greater rotational stability of fracture or osteotomy sites. However, there appears to be no experimental evidence that rotational stability is improved with intramedullary nail compression. This study evaluated the effect of intramedullary nail compression on cadaveric and composite tibial fragment rotation above and below a mid-tibial transverse osteotomy.
METHODS: Twelve composite and four matched pairs of human cadaveric tibiae were randomized into either a compression group or a non-compressed dynamized group. A bi-axial servo-hydraulic testing machine applied 5Nm of internal and external torque (2Nm/s) under constant axial loads of 375N and 750N. Rotation along each tibia and at the osteotomy site was recorded from the trajectories of infrared emitting diodes rigidly attached along the tibial shaft.
FINDINGS: In comparison to dynamized nails, intramedullary nail compression significantly reduced the rotation at the osteotomy site in both human (8.9 degrees reduction; P=0.007) and composite tibiae (2.5 degrees reduction; P=0.039) under 750N of axial load. A similar result was noted for simulated half-body weight loads (375N) for both composite (3.0 degrees reduction; P=0.009) and human tibiae (10.5 degrees reduction; P=0.003).
INTERPRETATION: Intramedullary nail compression more effectively reduced tibial fragment rotation about a mid-tibial osteotomy than conventional dynamized intramedullary nails because intramedullary nails created more osteotomy site compression than the application of body weight to tibiae instrumented with dynamized intramedullary nails.

Entities:  

Mesh:

Year:  2007        PMID: 17250938     DOI: 10.1016/j.clinbiomech.2006.11.009

Source DB:  PubMed          Journal:  Clin Biomech (Bristol, Avon)        ISSN: 0268-0033            Impact factor:   2.063


  3 in total

1.  The effect of different torque wrenches on rotational stiffness in compressive femoral nails: a biomechanical study.

Authors:  A A Karaarslan; N Acar
Journal:  Eur J Trauma Emerg Surg       Date:  2017-07-27       Impact factor: 3.693

2.  A novel nail providing more biomechanical rotational and axial stability than conventional interlocking nail in femur complex fracture model.

Authors:  A A Karaarslan; N Acar; Y Örgen; H Aycan; F Ertem; E Sesli
Journal:  Eur J Trauma Emerg Surg       Date:  2016-05-03       Impact factor: 3.693

3.  The functional results of tibial shaft fractures treated with intramedullary nail compressed by proximal tube.

Authors:  Ahmet Adnan Karaarslan; Nihat Acar; Hakan Aycan; Erhan Sesli
Journal:  Strategies Trauma Limb Reconstr       Date:  2016-02-02
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.