The real intraabdominal pressure during laparoscopy: comparison of different insufflators.

Volker R Jacobs1, John E Morrison.   

Abstract

STUDY
OBJECTIVE: To compare 5 laparoscopic insufflators with different gas flow rates with regard to accuracy of preset pressure setting versus real intraoperative intraabdominal pressure.
DESIGN: Prospective study (Canadian Task Force classification II-2).
SETTING: Fayette Medical Center, Alabama. PATIENTS: Five patients undergoing laparoscopic cholecystectomy.
INTERVENTIONS: Intraoperative intraabdominal and system pressure measurements during comparable laparoscopic procedures.
MEASUREMENTS AND MAIN RESULTS: Actual intraabdominal pressure was measured and compared between 5 different 10 to 20 L/min insufflators (Storz Laparoflator and Endoflator, Richard Wolf, BEI Medical and Snowden & Pencer) with a computer-based online data acquisition system. At a nominal pressure of 10 mm Hg, the mean intraabdominal pressure during the entire procedure was measured to be between 9.68 and 11.45 mm Hg. The mean intraabdominal pressure during laparoscopy for the insufflators showed a margin of error of 14.5%, with maximum intraabdominal pressure peaks measured between 14.65 and 17.87 mm Hg.
CONCLUSIONS: Within an error margin of <15% of the preset intraabdominal pressure setting, the insufflators evaluated can be considered pressure reliable. Although intraabdominal pressure peaks exceeding the nominal pressure temporarily reached up to 78.7% of the setting, no apparent complications were observed. Excessive pressure peaks seen in the previous reported laboratory model could not be confirmed during in vivo application.

Entities:  

Mesh:

Year:  2007        PMID: 17218239     DOI: 10.1016/j.jmig.2006.06.025

Source DB:  PubMed          Journal:  J Minim Invasive Gynecol        ISSN: 1553-4650            Impact factor:   4.137


  4 in total

1.  Benchtop evaluation of pressure barrier insufflator and standard insufflator systems.

Authors:  Kenneth G Nepple; Dorina Kallogjeri; Sam B Bhayani
Journal:  Surg Endosc       Date:  2012-07-26       Impact factor: 4.584

2.  Asystole following positive pressure insufflation of right pleural cavity: a case report.

Authors:  Kari M Forde-Thielen; Mojca R Konia
Journal:  J Med Case Rep       Date:  2011-06-30

3.  Laparoscopic management of obstructing small bowel GIST tumor.

Authors:  John E Morrison; Ian A Hodgdon
Journal:  JSLS       Date:  2013 Oct-Dec       Impact factor: 2.172

4.  Cardiorespiratory impact of intrathoracic pressure overshoot during artificial carbon dioxide pneumothorax: a randomized controlled study.

Authors:  Yunqin Ren; Xing Zhu; Hong Yan; Liyong Chen; Qingxiang Mao
Journal:  BMC Anesthesiol       Date:  2022-03-23       Impact factor: 2.217

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.