Literature DB >> 17218209

A 10-year comparison of explanted Hancock-II and Carpentier-Edwards supraannular bioprostheses.

Jagdish Butany1, Shaun W Leong, Kristopher S Cunningham, Geoffrey D'Cruz, Khenan Carmichael, Terrence M Yau.   

Abstract

BACKGROUND: Bioprosthetic heart valves are more frequently being used in valve replacement procedures today. Although second-generation bioprosthetic valves have improved functionality over their first-generation counterparts, they still often fail due to primary tissue degeneration.
METHODS: This study examines two second-generation porcine valves after surgical explantation, the Hancock-II (HAN; Medtronic Heart Valve Division, Irvine, CA, USA) and the Carpentier-Edwards supraannular (CE-SAV; Baxter Healthcare Corporation, now Edwards LifeSciences, Irvine, CA, USA), with special attention to morphological/histological changes and reasons for valve failure. A total of 98 HAN and 65 CE-SAV valves were explanted and seen over a 10-year period.
RESULTS: CE-SAV valves had a longer average implant duration than HAN valves (13.9+/-3.9 years vs. 10.0+/-5.1 years). Compared with HAN valves, CE-SAV valves also had a higher incidence of stent deformation (41.5% vs. 14.3%), calcification (75.4% vs. 54.1%), and pannus (100% vs. 91.8%).
CONCLUSIONS: The greater degenerative changes seen with CE-SAV valves over HAN valves may be due to the longer implant duration of CE-SAV valves in this series. To our knowledge, the present study is the first direct morphological comparison of these two valve models.

Entities:  

Mesh:

Year:  2007        PMID: 17218209     DOI: 10.1016/j.carpath.2006.06.003

Source DB:  PubMed          Journal:  Cardiovasc Pathol        ISSN: 1054-8807            Impact factor:   2.185


  2 in total

1.  Cross-linked xenogenic collagen implantation in the sheep model for vaginal surgery.

Authors:  Masayuki Endo; Iva Urbankova; Jaromir Vlacil; Siddarth Sengupta; Thomas Deprest; Bernd Klosterhalfen; Andrew Feola; Jan Deprest
Journal:  Gynecol Surg       Date:  2015-02-05

2.  Investigation of failure modes of explanted porcine valves in the mitral position.

Authors:  Kun Liu; Wentao Feng; Xianda Yang; Jinglun Shen; Haibo Zhang; Yubo Fan
Journal:  J Thorac Dis       Date:  2021-05       Impact factor: 2.895

  2 in total