Literature DB >> 17217832

[Clinical contrast of cervical pedicle screw fixation assisted by C-arm fluoroscopy or 3D navigation system].

Wei Tian1, Ya-jun Liu, Bo Liu, Qin Li, Lin Hu, Zhi-yu Li, Qiang Yuan.   

Abstract

OBJECTIVE: To evaluate the feasibility and accuracy of cervical (C(2)-C(7)) pedicle screw fixation assisted by C-arm fluoroscopy or 3D navigation system (CT-based navigation system or Iso-C 3D navigation system).
METHODS: One hundred and forty-five cervical pedicle screws inserted with C-arm fluoroscopy and 187 pedicle screws inserted with 3D navigation system were observed by postoperative CT or Iso-C 3D scan. The process of navigation was investigated.
RESULTS: In the 145 screws inserted with C-arm fluoroscopy, 133 screws are accurate (91.7%), and in the 187 screws inserted by 3D navigation system, 183 screws are accurate (97.9%) (chi2 = 6.705, P = 0.010), and the 4 misplaced screws occurred in the early stage of navigation system application. The process of navigation was investigated in twenty-five patients. With CT-based navigation, the mean time for registration and surface matching was 3.5 minutes; with Iso-C 3D navigation, the mean time for images collection and transfer was 6.2 minutes. The mean time for screw-marker insertion was 2 minutes.
CONCLUSIONS: 3D navigation system can increase accuracy of cervical pedicle screw fixation obviously.

Entities:  

Mesh:

Year:  2006        PMID: 17217832

Source DB:  PubMed          Journal:  Zhonghua Wai Ke Za Zhi        ISSN: 0529-5815


  1 in total

1.  CT evaluation of cervical pedicle in a Chinese population for surgical application of transpedicular screw placement.

Authors:  Zhu Ruofu; Yang Huilin; Hu Xiaoyun; He Xishun; Tang Tiansi; Chen Liang; Li Xigong
Journal:  Surg Radiol Anat       Date:  2008-03-20       Impact factor: 1.246

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.