[Radiologic evaluation of cup placement variation in conventional total hip arthroplasty].

U Leichtle1, N Gosselke, C J Wirth, M Rudert.   

Abstract

PURPOSE: Determination of the amount of variation in conventional acetabular cup positioning (radiological inclination and anteversion) in view of different factors that could influence the measured angles.
MATERIALS AND METHODS: The intended acetabular cup position of 45 degrees inclination and 20 degrees anteversion was checked radiologically in 950 patients who received a cementless total hip replacement. The qualifications of the surgeon, operated side and implant model were recorded and analyzed with respect to a possible correlation with the results. Since a "safe angle" cannot be defined without consideration of other variables, a tolerable deviation of the target position was investigated.
RESULTS: The mean inclination angle was 48.7 degrees (SD 7 degrees, minimum 28 degrees, maximum 75 degrees ). Anteversion was measured with a mean of 18.6 degrees (SD 9 degrees, minimum -9 degrees, maximum 50 degrees ). Assuming an acceptable deviation of +/- 5 degrees from the target position (45 degrees inclination and 20 degrees anteversion), only 22.7 % of the acetabular cups were in this range. In the case of an acceptable deviation of +/- 10 degrees, 34.5 % of the cups were still outside of the acceptable range. The qualifications of the surgeon, the implanted model, as well as the operated side did not have a significant influence on the results.
CONCLUSION: The common implantation technique yielded significant variation with respect to anteversion and inclination. The application of computer-aided navigation in the placement of acetabular cups would help to improve accuracy and reproducibility considerably in total hip arthroplasty.

Entities:  

Mesh:

Year:  2007        PMID: 17203443     DOI: 10.1055/s-2006-927085

Source DB:  PubMed          Journal:  Rofo        ISSN: 1438-9010


  4 in total

1.  Biomechanical model based evaluation of Total Hip Arthroplasty therapy outcome.

Authors:  Jörg Eschweiler; Julia Hawlitzky; Valentin Quack; Markus Tingart; Björn Rath
Journal:  J Orthop       Date:  2017-09-21

Review 2.  Technology in Arthroplasty: Are We Improving Value?

Authors:  Bradford S Waddell; Kaitlin Carroll; Seth Jerabek
Journal:  Curr Rev Musculoskelet Med       Date:  2017-09

3.  The John Charnley Award: risk factors for cup malpositioning: quality improvement through a joint registry at a tertiary hospital.

Authors:  Mark C Callanan; Bryan Jarrett; Charles R Bragdon; David Zurakowski; Harry E Rubash; Andrew A Freiberg; Henrik Malchau
Journal:  Clin Orthop Relat Res       Date:  2011-02       Impact factor: 4.176

Review 4.  Navigated cup implantation in hip arthroplasty.

Authors:  Johannes Beckmann; Dirk Stengel; Markus Tingart; Jürgen Götz; Joachim Grifka; Christian Lüring
Journal:  Acta Orthop       Date:  2009-10       Impact factor: 3.717