Inappropriate shock: a failure of SVT discriminators in a dual chamber ICD?

Amin Al-Ahmad1, Angela Tsiperfal, Henry H Hsia, Paul J Wang.   

Abstract

Inappropriate shock remains a major issue in patients with implantable cardioverter defibrillators. We examine an inappropriate shock delivered in a patient with atrial tachycardia that conducted 1:1. We reconstruct the device algorithms that led to therapy delivery and discuss programming changes that could be considered.

Entities:  

Mesh:

Year:  2006        PMID: 17201851     DOI: 10.1111/j.1540-8159.2006.00556.x

Source DB:  PubMed          Journal:  Pacing Clin Electrophysiol        ISSN: 0147-8389            Impact factor:   1.976