This chapter describes a number of aspects important for calibration of matrix-assisted laser desorption/ionization time-of-flight spectra prior to peptide mass fingerprinting searches. Both multipoint internal calibration and mass defect-based calibration is illustrated. The chapter describes how potential internal calibrants, like tryptic autodigest peptides and keratin-related peptides, can be identified and used for high-precision calibration. Furthermore, the construction of project/user-specific lists of potential calibrants is illustrated.

Mesh:

Substances:

Year:  2007        PMID: 17185769     DOI: 10.1385/1-59745-275-0:49

Source DB:  PubMed          Journal:  Methods Mol Biol        ISSN: 1064-3745


  2 in total

1.  An accurate and clean calibration method for MALDI-MS.

Authors:  Bosong Xiang; Mindy Prado
Journal:  J Biomol Tech       Date:  2010-09

2.  A mass spectrometry primer for mass spectrometry imaging.

Authors:  Stanislav S Rubakhin; Jonathan V Sweedler
Journal:  Methods Mol Biol       Date:  2010