Literature DB >> 17182711

An increased thermal index can be achieved when performing Doppler studies in obstetric sonography.

Eyal Sheiner1, Ilana Shoham-Vardi, Xavier Pombar, Michael J Hussey, Howard T Strassner, Jacques S Abramowicz.   

Abstract

OBJECTIVE: The present study was aimed at evaluating acoustic outputs during clinical ultrasound examinations, as expressed by the thermal index (TI) and the mechanical index (MI), during the second half of pregnancy and comparing acoustic outputs between B-mode and Doppler examinations.
METHODS: Patients with suspected fetal growth problems undergoing Doppler studies of the fetal circulation in addition to B-mode sonography were selected. Examinations took place between 21 and 40 weeks' gestation. An obstetrician collected data prospectively. Sonographers were unaware of the data being sought. The analysis of variance test was applied for differences in continuous variables.
RESULTS: A total of 63 examinations were evaluated. The mean gestational age +/- SD was 31.6 +/- 5.1 weeks. The mean duration of the total examinations was 17.6 +/- 8.6 minutes, whereas the Doppler studies lasted 0.9 +/- 0.8 minutes. The TI was significantly higher in the pulsed wave Doppler studies (mean, 1.5 +/- 0.5; range, 0.9-2.8) and color flow imaging studies (mean, 0.8 +/- 0.1; range, 0.6-1.2) compared with B-mode sonography (mean, 0.3 +/- 0.1; range, 0.1-0.7; P < .01). During the examination, 190 B-mode MI variations were recorded (mean, 1.1 +/- 0.1), which were comparable with those of the 31 color flow Doppler studies (mean, 1.0 +/- 0.1; P = .09) but higher than the 190 pulsed wave Doppler MI variations (mean 0.9 +/- 0.2; P < .001).
CONCLUSIONS: Increased acoustic output levels, as expressed by TI levels, are reached during obstetric Doppler studies. In particular, TI levels may reach 1.5 and higher. Doppler procedures should be performed with caution and be as brief as possible during obstetric sonography.

Entities:  

Mesh:

Year:  2007        PMID: 17182711     DOI: 10.7863/jum.2007.26.1.71

Source DB:  PubMed          Journal:  J Ultrasound Med        ISSN: 0278-4297            Impact factor:   2.153


  2 in total

1.  Do current Philips ultrasound systems exceed the recommended safety limits during routine prenatal ultrasounds?

Authors:  Qi Hao Ong; Martin Necas; Karen Lissington
Journal:  Australas J Ultrasound Med       Date:  2019-06-27

2.  Acoustic Power Measurement and Thermal Bioeffect Evaluation of Therapeutic Langevin Transducers.

Authors:  Jinhyuk Kim; Jungwoo Lee
Journal:  Sensors (Basel)       Date:  2022-01-14       Impact factor: 3.576