Mechanical torque measurement predicts load to implant cut-out: a biomechanical study investigating DHS anchorage in femoral heads.

Norbert Suhm1, Clemens Hengg, Ronald Schwyn, Markus Windolf, Volker Quarz, Markus Hänni.   

Abstract

INTRODUCTION: Bone strength plays an important role in implant anchorage. Bone mineral density (BMD) is used as surrogate parameter to quantify bone strength and to predict implant anchorage. BMD can be measured by means of quantitative computer tomography (QCT) or dual energy X-ray absorptiometry (DXA). These noninvasive methods for BMD measurement are not available pre- or intra-operatively. Instead, the surgeon could determine bone strength by direct mechanical measurement. We have evaluated mechanical torque measurement for (A) its capability to quantify local bone strength and (B) its predictive value towards load at implant cut-out.
MATERIALS AND METHODS: Our experimental study was performed using sixteen paired human cadaver proximal femurs. BMD was determined for all specimens by QCT. The torque to breakaway of the cancellous bone structure (peak torque) was measured by means of a mechanical probe at the exact position of subsequent DHS placement. The fixation strength of the DHS achieved was assessed by cyclic loading in a stepwise protocol beginning with 1,500 N increasing 500 N every 5,000 cycles until 4,000 N.
RESULTS: A highly significant correlation of peak torque with BMD (QCT) was found (r = 0.902, r (2) = 0.814, P < 0.001). Peak torque correlated highly significant with the load at implant cut-out (r = 0.795, P < 0.001). All specimens with a measured peak torque below 6.79 Nm failed at the first load level of 1,500 N. The specimens with a peak torque above 8.63 Nm survived until the last load level of 4,000 N.
CONCLUSION: Mechanical peak torque measurement is able to quantify bone strength. In an experimental setup, peak torque identifies those specimens that are likely to fail at low load. In clinical routine, implant migration and cut-out depend on several parameters, which are difficult to control, such as fracture type, fracture reduction achieved, and implant position. The predictive value of peak torque towards cut-out in a clinical set-up therefore has to be carefully validated.

Entities:  

Mesh:

Year:  2006        PMID: 17165032     DOI: 10.1007/s00402-006-0265-8

Source DB:  PubMed          Journal:  Arch Orthop Trauma Surg        ISSN: 0936-8051            Impact factor:   3.067


  12 in total

Review 1.  [Biomechanics of implant augmentation].

Authors:  M Windolf
Journal:  Unfallchirurg       Date:  2015-09       Impact factor: 1.000

2.  A novel method for intraoperative osseomechanical strength measurements: a biomechanical ex vivo evaluation on proximal femora.

Authors:  Alexander Maslaris; Olaf Brinkmann; Frank Layher; Georg Matziolis; Matthias Bungartz
Journal:  Arch Orthop Trauma Surg       Date:  2019-11-06       Impact factor: 3.067

3.  Local bone quality measure and construct failure prediction: a biomechanical study on distal femur fractures.

Authors:  Thomas Vordemvenne; Dirk Wähnert; Dominic Gehweiler; Ursula Styger; Boyko Gueorguiev; Christian Colcuc
Journal:  Arch Orthop Trauma Surg       Date:  2021-02-15       Impact factor: 2.928

4.  Mechanical torque measurement in the proximal femur correlates to failure load and bone mineral density ex vivo.

Authors:  Stefan Grote; Tatjana Noeldeke; Michael Blauth; Wolf Mutschler; Dominik Bürklein
Journal:  Orthop Rev (Pavia)       Date:  2013-06-24

5.  Utility of the cortical thickness of the distal radius as a predictor of distal-radius bone density.

Authors:  Sascha Rausch; Kajetan Klos; Florian Gras; Hristo Kostov Skulev; Albrecht Popp; Gunther Olaf Hofmann; Thomas Mückley
Journal:  Arch Trauma Res       Date:  2013-06-01

6.  Bone quality assessment for total hip arthroplasty with intraoperative trabecular torque measurements.

Authors:  Matthias C M Klotz; Nicholas A Beckmann; Rudi G Bitsch; Elisabeth Seebach; Tobias Reiner; Sebastian Jäger
Journal:  J Orthop Surg Res       Date:  2014-11-13       Impact factor: 2.359

7.  Implant augmentation: adding bone cement to improve the treatment of osteoporotic distal femur fractures: a biomechanical study using human cadaver bones.

Authors:  Dirk Wähnert; Ladina Hofmann-Fliri; R Geoff Richards; Boyko Gueorguiev; Michael J Raschke; Markus Windolf
Journal:  Medicine (Baltimore)       Date:  2014-11       Impact factor: 1.889

8.  Regional apparent density correlations within the proximal humerus.

Authors:  Jacob M Reeves; Tom Vanasse; Chris Roche; George S Athwal; James A Johnson; Kenneth J Faber; G Daniel G Langohr
Journal:  JSES Int       Date:  2021-02-09

9.  Mechanical quantification of local bone quality in the humeral head: a feasibility study.

Authors:  Alexander Scola; Florian Gebhard; Sebastian Weckbach; Christoph Dehner; Ronald Schwyn; Ladina Fliri; Götz Röderer
Journal:  Open Orthop J       Date:  2013-05-17

10.  Are radiographic indices reliable indicators for quantitative bone mineral density and vitamin D status after femoral neck fractures? A retrospective study in 112 elderly patients.

Authors:  Andy K S Yeo; Annette B Ahrberg; Jan D Theopold; Sebastian Ewens; Gudrun Borte; Christoph Josten; Johannes K M Fakler
Journal:  Patient Saf Surg       Date:  2015-11-25
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.