Activation processes in a medical linear accelerator and spatial distribution of activation products.

Helmut W Fischer1, Ben E Tabot, Björn Poppe.   

Abstract

Activation products have been identified by in situ gamma spectroscopy at the isocentre of a medical linear accelerator shortly after termination of a high energy photon beam irradiation with 15 x 15 cm field size. Spectra have been recorded either with an open or with a closed collimator. Whilst some activation products disappear from the spectrum with closed collimator or exhibit reduced count rates, others remain with identical intensity. The former isotopes are neutron-deficient and mostly decay by positron emission or electron capture; the latter have neutron excess and decay by beta(-) emission. This new finding is consistent with the assumption that photons in the primary beam produce activation products by (gamma, n) reactions in the treatment head and subsequently the neutrons created in these processes undergo (n, gamma) reactions creating activation products in a much larger area. These findings are expected to be generally applicable to all medical high energy linear accelerators.

Entities:  

Mesh:

Substances:

Year:  2006        PMID: 17148816     DOI: 10.1088/0031-9155/51/24/N02

Source DB:  PubMed          Journal:  Phys Med Biol        ISSN: 0031-9155            Impact factor:   3.609


  4 in total

1.  Secondary radiation dose during high-energy total body irradiation.

Authors:  M Janiszewska; K Polaczek-Grelik; M Raczkowski; B Szafron; A Konefał; W Zipper
Journal:  Strahlenther Onkol       Date:  2014-03-06       Impact factor: 3.621

2.  Thermal and resonance neutrons generated by various electron and X-ray therapeutic beams from medical linacs installed in polish oncological centers.

Authors:  Adam Konefał; Andrzej Orlef; Marcin Laciak; Aleksander Ciba; Marek Szewczuk
Journal:  Rep Pract Oncol Radiother       Date:  2012-07-12

3.  Technical Note: Time-gating to medical linear accelerator pulses: Stray radiation detector.

Authors:  Muhammad Ramish Ashraf; Petr Bruza; Venkat Krishnaswamy; David J Gladstone; Brian W Pogue
Journal:  Med Phys       Date:  2018-12-14       Impact factor: 4.071

4.  Cephalometric assessment of the axial inclination of upper and lower incisors in relation to the third-order angle.

Authors:  Michael Knösel; Rengin Attin; Dietmar Kubein-Meesenburg; Reza Sadat-Khonsari
Journal:  J Orofac Orthop       Date:  2007-05       Impact factor: 1.938

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.