Literature DB >> 17122971

Experimental results of mesh fixation by a manual manipulator in a laparoscopic inguinal hernia repair model.

N Inaki1, M Waseda, M O Schurr, M Braun, G F Buess.   

Abstract

BACKGROUND: Laparoscopic mesh fixation using a stapler can lead to complications such as nerve injury and bowel injury. However, mesh fixation by suturing with conventional laparoscopic instruments (CLI) is difficult because of limited degrees of freedom. A manual manipulator--Radius Surgical System (Radius)--whose tip can deflect and rotate, gives the surgeon two additional degrees of freedom. The aim of this study is to evaluate the introduction of Radius to mesh fixation in laparoscopic inguinal hernia repair.
METHODS: A model for inguinal hernia repair was prepared using animal organs in a trainer. Mesh fixation was performed using Radius, stapler, and CLI. Tensile strength during extraction of mesh toward the vertical direction, and execution time, were measured.
RESULTS: The mean number of fixation points of Radius, stapler, and CLI was 9.3 +/- 1.5, 8.5 +/- 1.4, and 9.0 +/- 1.0, respectively. The mean tensile strength of fixation of mesh of Radius, stapler, and CLI was 140.7 +/- 48.9, 73.1 +/- 23.4, and 53.6 +/- 31.5 (N), respectively. The mean tensile strength per one fixation point by Radius, stapler, and CLI was 16.5 +/- 5.3, 8.7 +/- 2.8, and 6.3 +/- 3.6 (N), respectively. The mean execution time of Radius, stapler, and CLI was 479 +/- 108, 54 +/- 31, and 431 +/- 77 (sec), respectively.
CONCLUSIONS: The mesh fixation by Radius was stronger than that by staples and CLI. Two additional degrees of freedom were useful in difficult angles. The introduction of Radius is feasible and facilitates the fixation of mesh with sutures in laparoscopic inguinal hernia repair.

Entities:  

Mesh:

Year:  2006        PMID: 17122971     DOI: 10.1007/s00464-006-0065-9

Source DB:  PubMed          Journal:  Surg Endosc        ISSN: 0930-2794            Impact factor:   4.584


  16 in total

1.  Treatment of inguinodynia after laparoscopic herniorrhaphy: a combined laparoscopic and fluoroscopic approach to the removal of helical tackers.

Authors:  J Wong; M Anvari
Journal:  Surg Laparosc Endosc Percutan Tech       Date:  2001-04       Impact factor: 1.719

2.  Prolonged abdominal wall pain caused by transfascial sutures used in the laparoscopic repair of incisional hernia.

Authors:  J Vermeulen; I Alwayn; L P S Stassen
Journal:  Surg Endosc       Date:  2003-06-19       Impact factor: 4.584

3.  An innovative trainer for surgical procedures using animal organs.

Authors:  M Waseda; N Inaki; L Mailaender; G F Buess
Journal:  Minim Invasive Ther Allied Technol       Date:  2005       Impact factor: 2.442

4.  Recurrent hernia following endoscopic total extraperitoneal repair.

Authors:  Pradeep K Chowbey; Samik Kumar Bandyopadhyay; Anil Sharma; Rajesh Khullar; Vandana Soni; Manish Baijal
Journal:  J Laparoendosc Adv Surg Tech A       Date:  2003-02       Impact factor: 1.878

5.  Evaluation of adhesion formation, mesh fixation strength, and hydroxyproline content after intraabdominal placement of polytetrafluoroethylene mesh secured using titanium spiral tacks, nitinol anchors, and polypropylene suture or polyglactin 910 suture.

Authors:  C S Joels; B D Matthews; K W Kercher; C Austin; H J Norton; T C Williams; B T Heniford
Journal:  Surg Endosc       Date:  2005-03-23       Impact factor: 4.584

6.  Outcomes of laparoscopic herniorrhaphy without fixation of mesh to the abdominal wall.

Authors:  Y S Khajanchee; D R Urbach; L L Swanstrom; P D Hansen
Journal:  Surg Endosc       Date:  2001-10       Impact factor: 4.584

7.  Laparoscopic total extraperitoneal hernia repair: mesh fixation is unnecessary.

Authors:  G C Beattie; S Kumar; S J Nixon
Journal:  J Laparoendosc Adv Surg Tech A       Date:  2000-04       Impact factor: 1.878

Review 8.  Mechanisms of hernia recurrence after preperitoneal mesh repair. Traditional and laparoscopic.

Authors:  A S Lowham; C J Filipi; R J Fitzgibbons; R Stoppa; G E Wantz; E L Felix; W B Crafton
Journal:  Ann Surg       Date:  1997-04       Impact factor: 12.969

9.  Tisseel versus tack staples as mesh fixation in totally extraperitoneal laparoscopic repair of groin hernias: a retrospective analysis.

Authors:  P Topart; F Vandenbroucke; P Lozac'h
Journal:  Surg Endosc       Date:  2005-03-11       Impact factor: 4.584

10.  Open mesh versus laparoscopic mesh repair of inguinal hernia.

Authors:  Leigh Neumayer; Anita Giobbie-Hurder; Olga Jonasson; Robert Fitzgibbons; Dorothy Dunlop; James Gibbs; Domenic Reda; William Henderson
Journal:  N Engl J Med       Date:  2004-04-25       Impact factor: 91.245

View more
  8 in total

1.  Single-incision laparoscopic hernioplasty with the assistance of the Radius Surgical System.

Authors:  Norihiko Ishikawa; Masahiko Kawaguchi; Satsuki Shimizu; Aika Matsunoki; Noriyuki Inaki; Go Watanabe
Journal:  Surg Endosc       Date:  2009-07-25       Impact factor: 4.584

2.  Laparoendoscopic single-site common bile duct exploration using the manual manipulator.

Authors:  Kazunori Shibao; Aiichiro Higure; Koji Yamaguchi
Journal:  Surg Endosc       Date:  2013-02-23       Impact factor: 4.584

3.  A meta-analysis of randomized controlled trials of fixation versus nonfixation of mesh in laparoscopic total extraperitoneal inguinal hernia repair.

Authors:  Yuan Jun Teng; Shu Mei Pan; Ya Li Liu; Ke Hu Yang; You Cheng Zhang; Jin Hui Tian; Jian Xu Han
Journal:  Surg Endosc       Date:  2011-04-13       Impact factor: 4.584

4.  Performance of Kymerax© precision-drive articulating surgical system compared to conventional laparoscopic instruments in a pelvitrainer model.

Authors:  Marco Alain Sieber; Bernhard Fellmann-Fischer; Michael Mueller
Journal:  Surg Endosc       Date:  2017-03-09       Impact factor: 4.584

5.  Laparoscopic Treatment for Esophageal Achalasia and Gastro-Esophago-reflex Disease Using Radius Surgical System.

Authors:  Yasumitsu Hirano; Noriyuki Inaki; Norihiko Ishikawa; Go Watanabe
Journal:  Indian J Surg       Date:  2012-06-16       Impact factor: 0.656

6.  Biomechanical properties of lightweight versus heavyweight meshes for laparoscopic inguinal hernia repair and their impact on recurrence rates.

Authors:  Christian Hollinsky; Simone Sandberg; Thomas Koch; Sabine Seidler
Journal:  Surg Endosc       Date:  2008-04-29       Impact factor: 4.584

Review 7.  Robot-like dexterity without computers and motors: a review of hand-held laparoscopic instruments with wrist-like tip articulation.

Authors:  Patrick L Anderson; Ray A Lathrop; Robert J Webster
Journal:  Expert Rev Med Devices       Date:  2016-07       Impact factor: 3.166

8.  Comparison of Laparoscopic Steerable Instruments Performed by Expert Surgeons and Novices.

Authors:  Luca Lacitignola; Rodrigo Trisciuzzi; Annarita Imperante; Laura Fracassi; Alberto Maria Crovace; Francesco Staffieri
Journal:  Vet Sci       Date:  2020-09-15
  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.