Accuracy and precision of a new portable ultrasound scanner, the BME-150A, in residual urine volume measurement: a comparison with the BladderScan BVI 3000.

Jin Ho Choe1, Ji Yeon Lee, Kyu-Sung Lee.   

Abstract

The objective of the study was to determine the relative accuracy of a new portable ultrasound unit, BME-150A, and the BladderScan BVI 3000, as assessed in comparison with the catheterized residual urine volume. We used both of these machines to prospectively measure the residual urine volumes of 89 patients (40 men and 49 women) who were undergoing urodynamic studies. The ultrasound measurements were compared with the post-scan bladder volumes obtained by catheterization in the same patients. The ultrasounds were followed immediately (within 5 min) by in-and-out catheterizations while the patients were in a supine position. There were a total of 116 paired measurements made. The BME-150A and the BVI 3000 demonstrated a correlation with the residual volume of 0.92 and 0.94, and a mean difference from the true residual volume of 7.8 and 3.6 ml, respectively. Intraclass correlation coefficients for the accuracy of the two bladder scans were 0.90 for BME-150A and 0.95 for BVI 3000. The difference of accuracy between the two models was not significant (p = 0.2421). There were six cases in which a follow-up evaluation of falsely elevated post-void residual urine volume measurements on the ultrasound studies resulted in comparatively low catheterized volumes, with a range of differences from 66 to 275.5 ml. These cases were diagnosed with an ovarian cyst, uterine myoma, or uterine adenomyosis on pelvic ultrasonography. The accuracy of the BME-150A is comparable to that of the BVI 3000 in estimating the true residual urine volumes and is sufficient enough for us to recommend its use as an alternative to catheterization.

Entities:  

Mesh:

Year:  2006        PMID: 17115233     DOI: 10.1007/s00192-006-0219-2

Source DB:  PubMed          Journal:  Int Urogynecol J Pelvic Floor Dysfunct


  13 in total

Review 1.  Cystic pelvic pathology presenting as falsely elevated post-void residual urine measured by portable ultrasound bladder scanning: report of 3 cases and review of the literature.

Authors:  M R Cooperberg; S K Chambers; T J Rutherford; H E Foster
Journal:  Urology       Date:  2000-04       Impact factor: 2.649

2.  Accuracy of total bladder volume and residual urine measurements: comparison between real-time ultrasonography and catheterization.

Authors:  T C Mainprize; H P Drutz
Journal:  Am J Obstet Gynecol       Date:  1989-04       Impact factor: 8.661

3.  The accuracy of portable abdominal ultrasound equipment in measuring postvoid residual volume.

Authors:  B Alnaif; H P Drutz
Journal:  Int Urogynecol J Pelvic Floor Dysfunct       Date:  1999

4.  Noninvasive bladder volume measurement.

Authors:  H Chan
Journal:  J Neurosci Nurs       Date:  1993-10       Impact factor: 1.230

5.  The accuracy of a hand-held real time ultrasound scanner for estimating bladder volume.

Authors:  G Ravichandran; G J Fellows
Journal:  Br J Urol       Date:  1983-02

6.  Ultrasonic studies of the bladder.

Authors:  J H Holmes
Journal:  J Urol       Date:  1967-04       Impact factor: 7.450

7.  Experience with portable ultrasound equipment and measurement of urine volumes: inter-user reliability and factors of patient position.

Authors:  T L Massagli; D D Cardenas; E W Kelly
Journal:  J Urol       Date:  1989-10       Impact factor: 7.450

8.  Application of portable ultrasound scanners in the measurement of post-void residual urine.

Authors:  Ching-Hwa Teng; Yu-Hui Huang; Bih-Jaw Kuo; Liu-Ing Bih
Journal:  J Nurs Res       Date:  2005-09       Impact factor: 1.682

9.  The accuracy of ultrasonic estimation of bladder volume: a comparison of portable and stationary equipment.

Authors:  Yu-Hui Huang; Liu-Ing Bih; Sung-Lang Chen; Su-Ju Tsai; Chin-Hwa Teng
Journal:  Arch Phys Med Rehabil       Date:  2004-01       Impact factor: 3.966

10.  Residual urine volumes in patients with spinal cord injury: measurement with a portable ultrasound instrument.

Authors:  D D Cardenas; E Kelly; J N Krieger; W H Chapman
Journal:  Arch Phys Med Rehabil       Date:  1988-07       Impact factor: 3.966

View more
  12 in total

1.  Ultrasound measurement of bladder wall thickness in different forms of detrusor overactivity.

Authors:  Maurizio Serati; Stefano Salvatore; Elena Cattoni; Marco Soligo; Antonella Cromi; Fabio Ghezzi
Journal:  Int Urogynecol J       Date:  2010-06-10       Impact factor: 2.894

2.  The accuracy of post-void residual measurement in women.

Authors:  Bernard T Haylen; Joseph Lee
Journal:  Int Urogynecol J Pelvic Floor Dysfunct       Date:  2008-02-27

3.  Accuracy and precision of a new portable ultrasound scanner, the Biocon-700, in residual urine volume measurement.

Authors:  Moon Kyoung Cho; Eun Ji Noh; Chul Hong Kim
Journal:  Int Urogynecol J       Date:  2016-12-10       Impact factor: 2.894

4.  Repeatability of post-void residual urine ≥ 100 ml in urogynaecologic patients.

Authors:  Marie-Louise Saaby; Gunnar Lose
Journal:  Int Urogynecol J       Date:  2011-12-13       Impact factor: 2.894

5.  Is submucosal bladder pressure monitoring feasible?

Authors:  Anisha S Basu; Steve Majerus; Elizabeth Ferry; Iryna Makovey; Hui Zhu; Margot S Damaser
Journal:  Proc Inst Mech Eng H       Date:  2018-01-29       Impact factor: 1.617

6.  [Association between residual urinary volume and urinary tract infection: prospective trial in 225 male patients].

Authors:  S Brookman-May; M Burger; B Hoschke; W F Wieland; F Kendel; C Gilfrich; K-P Braun; M May
Journal:  Urologe A       Date:  2010-09       Impact factor: 0.639

7.  Discrepancies in measuring bladder volumes with bedside ultrasound and bladder scanning in the intensive care unit: A pilot study.

Authors:  Donna M Prentice; Carrie Sona; Brian T Wessman; Enyo A Ablordeppey; Warren Isakow; Cassandra Arroyo; Marilyn Schallom
Journal:  J Intensive Care Soc       Date:  2017-11-13

8.  Frequency of recurrent urinary tract infection in patients with pelvic organ prolapse.

Authors:  Emrah Töz; Sefa Kurt; Çağdaş Sahin; Mehmet Tunç Canda
Journal:  Res Rep Urol       Date:  2015-01-28

9.  The accuracy of three-dimensional bladder ultrasonography in determining the residual urinary volume compared with conventional catheterisation.

Authors:  Imran K Jalbani; M Hammad Ather
Journal:  Arab J Urol       Date:  2014-07-29

10.  The reliability and reproducibility of ultrasonography for measuring the residual urine volume in men with lower urinary tract symptoms.

Authors:  Hassan A Abdelwahab; Housseini M Abdalla; Mahmoud H Sherief; Mohamed B Ibrahim; Mostafa A Shamaa
Journal:  Arab J Urol       Date:  2014-11-20
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.