AIM: To compare the shaping ability of Mtwo rotary nickel-titanium instruments in simulated curved canals and in curved canals of extracted teeth when set into permanent rotation with two different torque-limited automated devices.
METHODOLOGY: Root canal instrumentation was performed with two different torque-limited automated devices (Mtwo direct handpiece and Endo IT motor) using Mtwo rotary nickel-titanium instruments up to size 35. Simulated canals: 28 degrees and 35 degrees curved canals in resin blocks (n = 20 canals in each group) were prepared. Pre- and post-instrumentation images were recorded and assessment of canal shape was completed with a computer image analysis program. Extracted teeth: a total of 40 curved root canals were divided into two groups, which were balanced with respect to the angle and the radius of canal curvature. Straightening of the canal curvatures was determined with a computer image analysis program. Incidence of canal aberrations, preparation time, changes of working length and instrument failures were recorded both in simulated and real canals and were analysed using the t-test and a chi-square test.
RESULTS: In simulated canals, instrumentation with Endo IT was significantly faster than with Mtwo direct (P < 0.05). No significant differences between the two devices were noted when all other parameters were assessed (P > 0.05). Only one instrument separated (Mtwo direct) during the enlargement of 28 degrees curved simulated canals. All systems maintained working distance well.
CONCLUSIONS: Both systems respected original root canal curvature well and were safe, indicating that the Mtwo direct handpiece is suitable for preparing curved root canals with the Mtwo instruments.

Mesh:

Substances:

Year:  2006        PMID: 17107539     DOI: 10.1111/j.1365-2591.2006.01171.x

Source DB:  PubMed          Journal:  Int Endod J        ISSN: 0143-2885            Impact factor:   5.264


  9 in total

1.  Shaping ability of two nickel-titanium instruments activated by continuous rotation or adaptive motion: a micro-computed tomography study.

Authors:  E Pedullà; G Plotino; N M Grande; G Avarotti; G Gambarini; E Rapisarda; F Mannocci
Journal:  Clin Oral Investig       Date:  2016-02-15       Impact factor: 3.573

2.  Modifications in Canal Anatomy of Curved Canals of Mandibular First Molars by two Glide Path Instruments using CBCT.

Authors:  Anil Dhingra; Nayasha Manchanda
Journal:  J Clin Diagn Res       Date:  2014-11-20

3.  Comparison of two canal preparation techniques using mtwo rotary instruments.

Authors:  Faeze Hamze; Kiamars Honardar; Kiumars Nazarimoghadam
Journal:  Iran Endod J       Date:  2011-11-15

4.  Centering ability and dentin removal of rotary systems in curved root canals.

Authors:  Saeed Moradi; Ali Talati; Ali Monajem Zadeh
Journal:  Iran Endod J       Date:  2009-07-06

5.  Root canal shaping by single-file systems and rotary instruments: a laboratory study.

Authors:  Khaly Bane; Babacar Faye; Mouhamed Sarr; Seydina O Niang; Diouma Ndiaye; Pierre Machtou
Journal:  Iran Endod J       Date:  2015-03-18

6.  Ex Vivo Comparison of Mtwo and RaCe Rotary File Systems in Root Canal Deviation: One File Only versus the Conventional Method.

Authors:  Mohsen Aminsobhani; Hasan Razmi; Solmaz Nozari
Journal:  J Dent (Tehran)       Date:  2015-07

7.  Cleaning Effectiveness of a Reciprocating Single-file and a Conventional Rotary Instrumentation System.

Authors:  Fredson Marcio Acris de Carvalho; Leonardo Cantanhede de Oliveira Gonçalves; André Augusto Franco Marques; Vanessa Alves; Carlos Eduardo da Silveira Bueno; Alexandre Sigrist De Martin
Journal:  Open Dent J       Date:  2016-12-26

8.  Microcomputed tomographic evaluation of shaping ability of two thermo mechanically treated single-file systems in severely curved roots.

Authors:  Swapnil J Kolhe; Priyanka S Kolhe; Meenal N Gulve; Gayatri B Aher; Chetan J Bhadage; Shailendra S Mashalkar
Journal:  J Conserv Dent       Date:  2020-12-04

9.  Shaping ability of nickel-titanium rotary instruments in curved root canals.

Authors:  Ali Talati; Saeed Moradi; Maryam Forghani; Ali Monajemzadeh
Journal:  Iran Endod J       Date:  2013-05-01
  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.