The validity and reliability of intestinal temperature during intermittent running.

Nicholas Gant1, Greg Atkinson, Clyde Williams.   

Abstract

PURPOSE: In many sporting and occupational settings, measuring core temperature using invasive methods is not practical. In these circumstances, ingestible temperature sensor capsules are a promising alternative. To date, no researcher has investigated the validity of intestinal temperature (T(int)) during free running or examined the between-trial reproducibility of this technique. Therefore, in two investigations, we examined the validity and reliability of Tint during prolonged intermittent shuttle running.
METHODS: In investigation A, 10 male games players completed 60 min of exercise while their rectal temperature (T(rec)) and T(int) were monitored. In investigation B, Tint was measured while nine males undertook two 90-min bouts of exercise, separated by 7 d.
RESULTS: A mean systematic bias of -0.15 degrees C (95% CI, 0.10-0.20) was found between T(int) and T(rec) during exercise. This bias for T(int) to record higher temperatures than T(rec) was uniform through the range of measurements, such that the exercise-mediated changes in body temperature were similar between methods of measurement. The 95% limits of agreement were found to be +/- 0.22 degrees C (95% CI, 0.11-0.33) and correlations were high (r > 0.85), suggesting that random error between methods was acceptably small. In investigation B, the mean change between repeated trials was a negligible 0.01 degrees C (95% CI, -0.02 to 0.05). The within-subjects SD was 0.08 degrees C (95% CI, 0.05-0.15). Random error was uniform through the measurement range and was deemed acceptable on the basis of statistical power calculations.
CONCLUSION: These findings suggest that the test-retest variability of Tint is acceptably small during intermittent shuttle running. The small amount of random measurement error and similar thermal responses to exercise suggest that Tint is as appropriate for use in exercise physiology research as T(rec), provided that the consistent bias between these measurement methods is allowed for.

Mesh:

Year:  2006        PMID: 17095925     DOI: 10.1249/01.mss.0000233800.69776.ef

Source DB:  PubMed          Journal:  Med Sci Sports Exerc        ISSN: 0195-9131            Impact factor:   5.411


  26 in total

1.  The Big Picture of Individual Differences in Physical Activity Behavior Change: A Transdisciplinary Approach.

Authors:  Angela D Bryan; Renea Nilsson; Sara Anne Tompkins; Renee E Magnan; Bess H Marcus; Kent E Hutchison
Journal:  Psychol Sport Exerc       Date:  2011-01

2.  In a hot-dry environment racewalking increases the risk of hyperthermia in comparison to when running at a similar velocity.

Authors:  Ricardo Mora-Rodriguez; Juan F Ortega; Nassim Hamouti
Journal:  Eur J Appl Physiol       Date:  2010-11-27       Impact factor: 3.078

3.  Using an Ingestible Telemetric Temperature Pill to Assess Gastrointestinal Temperature During Exercise.

Authors:  Coen C W G Bongers; Maria T E Hopman; Thijs M H Eijsvogels
Journal:  J Vis Exp       Date:  2015-10-07       Impact factor: 1.355

Review 4.  The ingestible telemetric body core temperature sensor: a review of validity and exercise applications.

Authors:  Christopher Byrne; Chin Leong Lim
Journal:  Br J Sports Med       Date:  2006-12-18       Impact factor: 13.800

5.  Core body temperature during competition in the heat: National Boys' 14s Junior Championships.

Authors:  Michael F Bergeron; Kathryn S McLeod; John F Coyle
Journal:  Br J Sports Med       Date:  2007-06-11       Impact factor: 13.800

6.  Bioharness(™) multivariable monitoring device: part. I: validity.

Authors:  James A Johnstone; Paul A Ford; Gerwyn Hughes; Tim Watson; Andrew T Garrett
Journal:  J Sports Sci Med       Date:  2012-09-01       Impact factor: 2.988

7.  Bioharness(™) Multivariable Monitoring Device: Part. II: Reliability.

Authors:  James A Johnstone; Paul A Ford; Gerwyn Hughes; Tim Watson; Andrew T Garrett
Journal:  J Sports Sci Med       Date:  2012-09-01       Impact factor: 2.988

8.  Comparison of rectal and aural core body temperature thermometry in hyperthermic, exercising individuals: a meta-analysis.

Authors:  Robert Huggins; Neal Glaviano; Naoki Negishi; Douglas J Casa; Jay Hertel
Journal:  J Athl Train       Date:  2012 May-Jun       Impact factor: 2.860

9.  Pulmonary artery and intestinal temperatures during heat stress and cooling.

Authors:  James Pearson; Matthew S Ganio; Thomas Seifert; Morten Overgaard; Niels H Secher; Craig G Crandall
Journal:  Med Sci Sports Exerc       Date:  2012-05       Impact factor: 5.411

10.  Validity and reliability of devices that assess body temperature during indoor exercise in the heat.

Authors:  Matthew S Ganio; Christopher M Brown; Douglas J Casa; Shannon M Becker; Susan W Yeargin; Brendon P McDermott; Lindsay M Boots; Paul W Boyd; Lawrence E Armstrong; Carl M Maresh
Journal:  J Athl Train       Date:  2009 Mar-Apr       Impact factor: 2.860

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.