Literature DB >> 17087653

Analytical validation of the new plasma calibrated Accu-Chek Test Strips (Roche Diagnostics).

Cécile Meex1, José Poncin, Jean-Paul Chapelle, Etienne Cavalier.   

Abstract

BACKGROUND: The Accu-Chek Inform glucose monitor is a point-of-care system for testing blood glucose. New test strips, calibrated to deliver glucose plasma-like values, were launched on the market in May 2005. The aim of our study was to perform analytical validation of these new strips.
METHODS: We compared the new plasma strips with whole blood strips; results for the plasma strips with plasma values obtained using a clinical analyzer and with whole blood values given by the glucose electrode of a blood gas analyzer; and the influence of the type of blood (capillary or venous) on the results obtained by the glucose monitor with the plasma calibrated strips.
RESULTS: Plasma strips give on average 7% higher results than the previous whole blood strips. However, the results given by the plasma strips on capillary whole blood, even if well correlated, are not completely comparable with those given by an analyzer for venous plasma. Nevertheless, these plasma strips and the glucose electrode of a blood gas analyzer give comparable results.
CONCLUSIONS: Accu-Chek Inform plasma strips are a good method for monitoring of blood glucose values in patients with diabetes.

Entities:  

Mesh:

Substances:

Year:  2006        PMID: 17087653     DOI: 10.1515/CCLM.2006.249

Source DB:  PubMed          Journal:  Clin Chem Lab Med        ISSN: 1434-6621            Impact factor:   3.694


  4 in total

1.  Galactokinase deficiency in a patient with congenital hyperinsulinism.

Authors:  Mashbat Bayarchimeg; Dunia Ismail; Amanda Lam; Derek Burk; Jeremy Kirk; Wolfgang Hogler; Sarah E Flanagan; Sian Ellard; Khalid Hussain
Journal:  JIMD Rep       Date:  2011-12-13

2.  Accuracy of bedside glucometry in critically ill children with peripheral hypoperfusion.

Authors:  Anant Khositseth; Onsuthi Pharadornuwat; Rojjanee Lertbunrian; Nattachai Anantasit; Jarin Vaewpanich
Journal:  J Clin Monit Comput       Date:  2014-05-07       Impact factor: 2.502

3.  Inaccuracy of Venous Point-of-Care Glucose Measurements in Critically Ill Patients: A Cross-Sectional Study.

Authors:  Adriano José Pereira; Thiago Domingos Corrêa; Francisca Pereira de Almeida; Rodrigo Octávio Deliberato; Michelle dos Santos Lobato; Nelson Akamine; Eliézer Silva; Alexandre Biasi Cavalcanti
Journal:  PLoS One       Date:  2015-06-12       Impact factor: 3.240

4.  Effect of Dietary Inclusion of Gynura divaricata (L.) on Growth Performance, Hematology, and Carcass Fat Deposition in Broilers.

Authors:  Chairat Jamjang; Suwanna Kijpakorn; Kris Angkanaporn
Journal:  J Poult Sci       Date:  2020-04-25       Impact factor: 1.425

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.