Automated analysis of pleural fluid total and differential leukocyte counts with the Sysmex XE-2100.

Robert de Jonge1, Rob Brouwer, Marcel van Rijn, Bernadette A C van Acker, Henk J A M Otten, Jan Lindemans.   

Abstract

BACKGROUND: Determination of leukocyte (WBC) counts in pleural fluid is routinely performed by microscopic examination. In this study, we evaluated the performance of automated (differential) WBC counting in comparison with manual counting.
METHODS: Pleural fluid samples (n=45) were obtained from patients undergoing diagnostic thoracocentesis. The manual total WBC count was determined after Samson staining in a Fuchs-Rosenthal hemocytometer; microscopic differential counts were performed on May-Grünwald Giemsa-stained cytospin slides. The Sysmex XE-2100 hematology analyzer was used for automated (differential) WBC counting. The functional detection limit was determined by serial dilution of continuous ambulatory peritoneal dialysis (CAPD) fluid and replicate measurements of each dilution.
RESULTS: The automated WBC count (x10(6)/L) was highly correlated with that of the microscopic reference method (r(2)=0.95; WBC-analyzer=0.97 x WBC-reference method+16; n=45). Good agreement was also observed for the absolute lymphocyte count (r(2)=0.92; WBC-analyzer=0.99 x WBC-reference method+32; n=36), neutrophil count (r(2)=0.94; WBC-analyzer=0.91 x WBC-reference method+6; n=35), and monocyte count (r(2)=0.73; WBC-analyzer=0.83 x WBC-reference method+6; n=38). The functional detection limit for WBCs was calculated at 50 x 10(6)/L (coefficient of variation 20%).
CONCLUSIONS: With some limitations, total and differential WBC counts in pleural fluid can be reliably determined using the Sysmex XE-2100 instrument.

Entities:  

Mesh:

Year:  2006        PMID: 17087651     DOI: 10.1515/CCLM.2006.242

Source DB:  PubMed          Journal:  Clin Chem Lab Med        ISSN: 1434-6621            Impact factor:   3.694


  4 in total

1.  Continuous ambulatory peritoneal dialysis, ascitic and pleural body fluids evaluation with the Mindray BC-6800 hematology analyzer.

Authors:  Oscar Fuster; Belinda Andino; Amparo Pardo; Begoña Laiz
Journal:  J Clin Lab Anal       Date:  2017-04-26       Impact factor: 2.352

2.  Automated analysis for differentiating leukocytes in body fluids using the software "biological liquid application" on ADVIA2120/2120i hematology analyzer.

Authors:  Alessandra Falda; Paolo Doretto
Journal:  J Clin Lab Anal       Date:  2018-05-25       Impact factor: 2.352

3.  A flow cytometric method for determination of the blood neutrophil fraction in rats.

Authors:  Spela Skrajnar; Marija Anzur Lasnik; Apolonija Bedina Zavec
Journal:  J Am Assoc Lab Anim Sci       Date:  2009-03       Impact factor: 1.232

4.  Analytical validation of the Sysmex XT-2000iV for cell counts in canine and feline effusions and concordance with cytologic diagnosis.

Authors:  Nazaré Pinto da Cunha; Alessia Giordano; Mario Caniatti; Saverio Paltrinieri
Journal:  Vet Clin Pathol       Date:  2009-02-02       Impact factor: 1.180

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.