Literature DB >> 17085168

Long-term mechanical reliability of AMS 700CX/CXM inflatable penile prosthesis.

Nivedita Bhatta Dhar1, Kenneth W Angermeier, Drogo K Montague.   

Abstract

PURPOSE: We determined the long-term mechanical reliability of the AMS 700CX/CXM inflatable penile prosthesis in men with organic erectile dysfunction.
MATERIALS AND METHODS: A total of 455 consecutive patients underwent implantation of an AMS 700CX/CXM inflatable penile prosthesis between April 1986 and July 2004. Using telephone interviews and mailed questionnaires we obtained followup on 380 of 455 patients (83%). Data were obtained with regard to 2 end points, including overall failure and mechanical failure caused by any malfunctioning component.
RESULTS: Followup was 0.49 to 231 months (median 91.5). Ten-year Kaplan-Meier estimates of overall and mechanical survival were 74.9% (95% CI 69.2 to 81.1) and 81.3% (95% CI 75.7 to 87.3), respectively.
CONCLUSIONS: To our knowledge this is the longest followup of any inflatable penile prosthesis showing overall and mechanical survival by Kaplan-Meier estimates.

Entities:  

Mesh:

Year:  2006        PMID: 17085168     DOI: 10.1016/j.juro.2006.08.012

Source DB:  PubMed          Journal:  J Urol        ISSN: 0022-5347            Impact factor:   7.450


  14 in total

Review 1.  A practical overview of considerations for penile prosthesis placement.

Authors:  Landon Trost; Philip Wanzek; George Bailey
Journal:  Nat Rev Urol       Date:  2015-12-01       Impact factor: 14.432

Review 2.  [Surgical treatment of erectile dysfunction].

Authors:  T S Pottek
Journal:  Urologe A       Date:  2015-05       Impact factor: 0.639

Review 3.  Dual implantation of penile prosthesis and ATOMS® system for post-prostatectomy erectile dysfunction and urinary incontinence: a feasibility study.

Authors:  Marco Falcone; Mirko Preto; Enrico Ammirati; Gideon Blecher; Roberto Carone; Paolo Gontero; Alessandro Giammò
Journal:  Int J Impot Res       Date:  2020-06-04       Impact factor: 2.896

4.  Penile prosthesis implantation for the treatment for male erectile dysfunction: clinical outcomes and lessons learnt after 955 procedures.

Authors:  E Chung; C T Van; I Wilson; R A Cartmill
Journal:  World J Urol       Date:  2012-03-29       Impact factor: 4.226

Review 5.  Current management of penile implant infections, device reliability, and optimizing cosmetic outcome.

Authors:  John J Mulcahy; Andrew Kramer; William O Brant; Justin L Parker; Paul E Perito; Jeremy B Myers; Richard Bryson; Meagan Dunne
Journal:  Curr Urol Rep       Date:  2014-06       Impact factor: 3.092

Review 6.  Erectile dysfunction.

Authors:  Faysal A Yafi; Lawrence Jenkins; Maarten Albersen; Giovanni Corona; Andrea M Isidori; Shari Goldfarb; Mario Maggi; Christian J Nelson; Sharon Parish; Andrea Salonia; Ronny Tan; John P Mulhall; Wayne J G Hellstrom
Journal:  Nat Rev Dis Primers       Date:  2016-02-04       Impact factor: 52.329

Review 7.  Toward a new 'EPOCH': optimising treatment outcomes with phosphodiesterase type 5 inhibitors for erectile dysfunction.

Authors:  R Sadovsky; G B Brock; S W Gutkin; S Sorsaburu
Journal:  Int J Clin Pract       Date:  2009-08       Impact factor: 2.503

Review 8.  Narrative review of penile prosthetic implant technology and surgical results, including transgender patients.

Authors:  Michael Polchert; Brian Dick; Omer Raheem
Journal:  Transl Androl Urol       Date:  2021-06

9.  Diagnosis and treatment of erectile dysfunction--a practical update.

Authors:  C Persu; V Cauni; S Gutue; Elena Simona Albu; V Jinga; P Geavlete
Journal:  J Med Life       Date:  2009 Oct-Dec

Review 10.  Penile prosthesis surgery in the management of erectile dysfunction.

Authors:  Hossein Sadeghi-Nejad; Mina Fam
Journal:  Arab J Urol       Date:  2013-07-02
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.