Accuracy of dental digitizers.

Simon T Vlaar1, Jef M van der Zel.   

Abstract

UNLABELLED: The need for proper validation and verification methodology for CAD/CAM systems is imminent. CAD/CAM systems consisting of an optical impression system, design software and a fabrication machine have to perform to a certain level, whereby manufacturers need to prove the effectiveness of the system as a whole. However, especially when dental surface digitization devices are used as open, stand-alone applications in dental outsourcing, a reliable standard test for comparison is necessary.
PURPOSE: This study evaluates a proposed test method to be used to quantify "digitizing quality" with respect to accuracy and reproducibility of two dental surface digitization devices. Comparability of the characteristics should be ensured.
METHOD: Two laser light section scanners: DentaScope II and D200 were evaluated by means of the 'Sphere Test', that involves repeated measurements of a precision ball (radius: 6mm) according to a pre-defined protocol. The surface information was received as unmatched, overlapping point clouds and statistically processed with a newly developed software package. The standard deviation of all points as well as a measure for undercutting the equator were determined.
RESULTS: The standard deviation for the radius for D and S were 7.7 (+/-0.8) and 13.9 (+/-1.0) microm respectively. The equator undercut elevations were -2.0 degrees and -0.25 degrees for scanner D and S respectively.
CONCLUSION: Scanner D had a significantly higher accuracy than S (p<0.05), corresponding with the smaller pixel distance of the sensor. Both devices show adequate accuracy and reproducibility and have an adequate ability to detect the equator. The test is also suitable for calibration purposes.

Entities:  

Mesh:

Year:  2006        PMID: 17069074     DOI: 10.1111/j.1875-595x.2006.tb00105.x

Source DB:  PubMed          Journal:  Int Dent J        ISSN: 0020-6539            Impact factor:   2.512


  4 in total

1.  Clinical evaluation comparing the fit of all-ceramic crowns obtained from silicone and digital intraoral impressions.

Authors:  Cristina Zarauz; Arelhys Valverde; Francisco Martinez-Rus; Bassam Hassan; Guillermo Pradies
Journal:  Clin Oral Investig       Date:  2015-09-12       Impact factor: 3.573

2.  Accuracy in dental medicine, a new way to measure trueness and precision.

Authors:  Andreas Ender; Albert Mehl
Journal:  J Vis Exp       Date:  2014-04-29       Impact factor: 1.355

3.  Evaluating the accuracy of three intraoral scanners using models containing different numbers of crown-prepared abutments.

Authors:  Ting Zhang; Ting Wei; Yawen Zhao; Mengyang Jiang; Xiaojie Yin; Huiqiang Sun
Journal:  J Dent Sci       Date:  2021-06-01       Impact factor: 2.080

4.  Digital Dental Models: Is Photogrammetry an Alternative to Dental Extraoral and Intraoral Scanners?

Authors:  Francesca Zotti; Luca Rosolin; Massimo Bersani; Andrea Poscolere; Davide Pappalardo; Nicoletta Zerman
Journal:  Dent J (Basel)       Date:  2022-02-07
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.