| Literature DB >> 17034004 |
S Chowdhury1, Yogesh K Vohra, Jack E Lemons, Masaru Ueno, Junji Ikeda.
Abstract
Recently, alternations of zirconia ceramic femoral heads of total hip prostheses during in vivo conditions have caused concern in the medical disciplines regarding phase transformation of zirconia prosthetic components. In this paper, we have investigated the mechanical and structural properties of different laboratory aged zirconia femoral heads and correlated changes in mechanical properties with the phase compositions of the sample. From laser microscope observation, cross-sectional Scanning electron microscopy imaging, and X-ray diffraction analysis on the surface of the zirconia femoral heads, we found monoclinic to tetragonal phase transformation in zirconia prostheses over time during the aging process in the laboratory. Mechanical properties, mainly hardness (H) and Young's modulus (E) values, were measured by nanoindentation technique on the surface of these implants. The results showed that both H and E values decreased with increased monoclinic phase in zirconia, thus confirming a phase transformation over time during aging.Entities:
Mesh:
Substances:
Year: 2007 PMID: 17034004 DOI: 10.1002/jbm.b.30688
Source DB: PubMed Journal: J Biomed Mater Res B Appl Biomater ISSN: 1552-4973 Impact factor: 3.368