Thermal-mechanical study of functionally graded dental implants with the finite element method.

F Wang1, H P Lee, C Lu.   

Abstract

This article investigates the thermal-mechanical performance of hydroxyapatite/titanium (HA/Ti) functionally graded (FG) dental implants with the three-dimensional finite element method. The stresses induced by occlusal force for the present HA/Ti FG implant are calculated to compare with the corresponding stresses for the titanium dental implant. Thermal-mechanical effect of temperature variation due to daily oral activity is also studied. The HA/Ti FG dental implant performance is evaluated against the maximum von Mises stress, which is the general performance indicator, the first principal/tensile stress for mechanical failure of implant-bone-bond and the third principal/compressive stress for bone absorption. Simulation results indicate that under the influence of occlusal force only, the FG implants with different HA fraction along the implant length perform almost equally well, while the titanium implant sustains much higher von Mises stress. However, when thermal stress is also considered, the FG implant having HA fraction exponential index of m = 2 with temperature decrease of 20 degrees C yields the highest first principal and von Mises stresses among all the FG and titanium implants. (c) 2006 Wiley Periodicals, Inc.

Entities:  

Mesh:

Substances:

Year:  2007        PMID: 17001649     DOI: 10.1002/jbm.a.30855

Source DB:  PubMed          Journal:  J Biomed Mater Res A        ISSN: 1549-3296            Impact factor:   4.396


  6 in total

Review 1.  Functional Gradient Metallic Biomaterials: Techniques, Current Scenery, and Future Prospects in the Biomedical Field.

Authors:  Hongyuan Shi; Peng Zhou; Jie Li; Chaozong Liu; Liqiang Wang
Journal:  Front Bioeng Biotechnol       Date:  2021-01-18

2.  Investigation of the effects of graded models on the biomechanical behavior of a bone-dental implant system under osteoporotic conditions.

Authors:  Ying Li; Zhong Shuang Liu; Xiao Ming Bai; Bin Zhang
Journal:  Pak J Med Sci       Date:  2013-04       Impact factor: 1.088

Review 3.  Dental prostheses mimic the natural enamel behavior under functional loading: A review article.

Authors:  Ahmed A Madfa; Xiao-Guang Yue
Journal:  Jpn Dent Sci Rev       Date:  2015-09-01

4.  Application of calcium phosphate materials in dentistry.

Authors:  Jabr S Al-Sanabani; Ahmed A Madfa; Fadhel A Al-Sanabani
Journal:  Int J Biomater       Date:  2013-06-26

5.  Comparison of various functionally graded femoral prostheses by finite element analysis.

Authors:  Azim Ataollahi Oshkour; Hossein Talebi; Seyed Farid Seyed Shirazi; Mehdi Bayat; Yat Huang Yau; Faris Tarlochan; Noor Azuan Abu Osman
Journal:  ScientificWorldJournal       Date:  2014-08-27

6.  Thermal Load and Heat Transfer in Dental Titanium Implants: An Ex Vivo-Based Exact Analytical/Numerical Solution to the 'Heat Equation'.

Authors:  Grigorios P Panotopoulos; Ziyad S Haidar
Journal:  Dent J (Basel)       Date:  2022-03-10
  6 in total