Usefulness of a novel ultrasound transducer for continuous monitoring treadmill exercise echocardiography to assess coronary artery disease.

Kenichi Nakashiki1, Akira Kisanuki, Yutaka Otsuji, Shiro Yoshifuku, Toshinori Yuasa, Kunitsugu Takasaki, Eiji Kuwahara, Bo Yu, Takeshi Uemura, Naoko Mizukami, Shuichi Hamasaki, Shinichi Minagoe, Chuwa Tei.   

Abstract

BACKGROUND: The feasibility of a novel ultrasound probe, which can be attached to the left ventricular (LV) apex chest wall and allows free rotation around its long axis direction for the continuous monitoring of LV wall motion, was tested. METHODS AND
RESULTS: There were 36 subjects who had coronary artery disease (CAD). By attaching a novel ultrasound probe to the chest wall, the LV apical views were recorded during treadmill exercise stress echocardiography (Echo). The continuous monitoring of LV wall motion was satisfactorily feasible in 30 of 36 patients. The visualization rate of the overall LV segments was higher at rest (90%) compared to that during peak exercise (77%). The segments were better visualized in apical portions (90-100%) than in mid (77-96%) or basal portions (68-87%). The sensitivity, specificity, and accuracy for detecting CAD were 61, 100 and 77%, respectively. The wall motion score index 3 and 6 min after exercise decreased significantly compared to those at peak exercise. The number of segments with dyssynergy was highest at the peak exercise. Ischemic ST-T depression on electrocardiography was observed only at peak stress periods.
CONCLUSIONS: Continuous monitoring treadmill exercise Echo using a novel ultrasound probe seems feasible for the non-invasive and physiological assessment of CAD.

Entities:  

Mesh:

Year:  2006        PMID: 16998262     DOI: 10.1253/circj.70.1297

Source DB:  PubMed          Journal:  Circ J        ISSN: 1346-9843            Impact factor:   2.993


  2 in total

1.  Image acquisition stability of fixated musculoskeletal sonography in an exercise setting: a quantitative analysis and comparison with freehand acquisition.

Authors:  H Maarten Heres; Marloes Sjoerdsma; Thijs Schoots; Marcel C M Rutten; Frans N van de Vosse; Richard G P Lopata
Journal:  J Med Ultrason (2001)       Date:  2019-11-07       Impact factor: 1.314

2.  A demonstration of high field-of-view stability in hands-free echocardiography.

Authors:  Marloes Sjoerdsma; Louis S Fixsen; Thijs Schoots; Frans N van de Vosse; Richard Gp Lopata
Journal:  Cardiovasc Ultrasound       Date:  2020-05-29       Impact factor: 2.062

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.