Assessing the difference between planned and delivered intensity-modulated radiotherapy dose distributions based on radiobiological measures.

P Mavroidis1, B C Ferreira, N Papanikotaou, R Svensson, C Kappas, B K Lind, A Brahme.   

Abstract

AIMS: Because of the highly conformal distributions that can be obtained with intensity-modulated radiotherapy (IMRT), any discrepancy between the intended and delivered distributions would probably affect the clinical outcome. Consequently, there is a need for a measure that would quantify those differences in terms of a change in the expected clinical outcome.
MATERIALS AND METHODS: To evaluate such a measure, cancer of the cervix was used, where the bladder and rectum are proximal and partially overlapping with the internal target volume. A solid phantom simulating the pelvic anatomy was fabricated and a treatment plan was developed to deliver the prescribed dose to the phantom. The phantom was then irradiated with films positioned in several transverse planes. The racetrack microtron at 50 MV was used in the treatment planning and delivery processes. The dose distribution delivered was analysed based on the film measurements and compared against the treatment plan. The differences in the measurements were evaluated using both physical and biological criteria. Whereas the physical comparison of dose distributions can assess the geometric accuracy of delivery, it does not reflect the clinical effect of any measured dose discrepancies.
RESULTS: It is shown how small inaccuracies in delivered dose can affect the treatment outcome in terms of complication-free tumour cure.
CONCLUSIONS: With highly conformal IMRT, the accuracy of the patient set-up and treatment delivery are critical for the success of the treatment. A method is proposed to evaluate the precision of the delivered plan based on changes in complication and control rates as they relate to uncertainties in dose delivery.

Entities:  

Mesh:

Year:  2006        PMID: 16969983     DOI: 10.1016/j.clon.2006.04.007

Source DB:  PubMed          Journal:  Clin Oncol (R Coll Radiol)        ISSN: 0936-6555            Impact factor:   4.126


  4 in total

1.  Evaluation on lung cancer patients' adaptive planning of TomoTherapy utilising radiobiological measures and Planned Adaptive module.

Authors:  Fan-Chi Su; Chengyu Shi; Panayiotis Mavroidis; Prema Rassiah-Szegedi; Niko Papanikolaou
Journal:  J Radiother Pract       Date:  2009-01-01

2.  Assessing four-dimensional radiotherapy planning and respiratory motion-induced dose difference based on biologically effective uniform dose.

Authors:  F-C Su; C Shi; P Mavroidis; V Goytia; R Crownover; P Rassiah-Szegedi; N Papanikolaou
Journal:  Technol Cancer Res Treat       Date:  2009-06

3.  Radiobiological evaluation of forward and inverse IMRT using different fractionations for head and neck tumours.

Authors:  Brigida C Ferreira; Maria do Carmo Lopes; Josefina Mateus; Miguel Capela; Panayiotis Mavroidis
Journal:  Radiat Oncol       Date:  2010-06-22       Impact factor: 3.481

4.  Evaluation of a commercial DIR platform for contour propagation in prostate cancer patients treated with IMRT/VMAT.

Authors:  Jacob E Hammers; Sara Pirozzi; Daniel Lindsay; Orit Kaidar-Person; Xianming Tan; Ronald C Chen; Shiva K Das; Panayiotis Mavroidis
Journal:  J Appl Clin Med Phys       Date:  2020-02       Impact factor: 2.102

  4 in total