Dosimetric verification of 6 and 18 MV intensity modulated photon beams using a dedicated fluoroscopic electronic portal imaging device (EPID).

Angelo F Monti1, Giovanna Frigerio.   

Abstract

BACKGROUND AND
PURPOSE: Intensity modulated radiation therapy offers a dose distribution improvement by modulating the two-dimensional X-ray fluence. This increases the treatment complexity from the planning phase to the verification one. Pre-treatment dosimetric verifications of each treatment field are strictly necessary, films and EPIDs are generally used for this purpose. In this paper we investigated the dosimetric characteristics of a commercially available fluoroscopic EPID "I'mRT QA" (Scanditronix-Wellhöfer) (I-QA) based on charge-coupled device camera (CCD) and we present a new method for optimising the EPID's response.
MATERIALS AND METHODS: I-QA is an optically sealed EPID and it is designed for on-line two-dimensional measurements of relative dose distribution. Dose profiles measured in water with diodes and dose distributions measured in water-equivalent phantoms with films were compared with those obtained with the I-QA for homogeneous and intensity modulated 6 and 18 MV photon beams. RESULTS AND
CONCLUSIONS: I-QA measurements depend on the field size and on the two-dimensional energy spectrum of the beam. The incoming beam was modified positioning a series of lead and plastic (RW3) slabs above the fluorescent screen to obtain a homogeneous response of the I-QA over the whole sensitive area. The thickness of lead and RW3 sheets was optimised to get the best matching between diodes, films and the I-QA measurements for each energy. Gamma index evaluations showed a correspondence between I-QA measurements and diode ones within 3% or 2 mm for homogeneous and simple modulated fields, and within 5% or 3mm for complex modulated fields.

Entities:  

Mesh:

Year:  2006        PMID: 16962676     DOI: 10.1016/j.radonc.2006.08.008

Source DB:  PubMed          Journal:  Radiother Oncol        ISSN: 0167-8140            Impact factor:   6.280


  1 in total

1.  Clinical examples of 3D dose distribution reconstruction, based on the actual MLC leaves movement, for dynamic treatment techniques.

Authors:  Wojciech Osewski; Lukasz Dolla; Michał Radwan; Marta Szlag; Roman Rutkowski; Barbara Smolińska; Krzysztof Slosarek
Journal:  Rep Pract Oncol Radiother       Date:  2014-07-17
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.