An in vitro evaluation of titanium, zirconia, and alumina procera abutments with hexagonal connection.

Paolo Vigolo1, Fulvio Fonzi, Zeina Majzoub, Giampiero Cordioli.   

Abstract

PURPOSE: The purpose of this study was to assess the precision at the implant interface of titanium, zirconia, and alumina Procera abutments with a hexagonal connection for single-tooth restorations.
MATERIALS AND METHODS: Twenty Procera abutments were produced with commercially pure titanium, 20 with zirconia, and 20 with alumina using computer-assisted design and manufacture (CAD/CAM). The rotational freedom of the abutments was assessed to detect the precision of fit of each abutment on the top of the implant hexagon.
RESULTS: Significant differences relative to rotational freedom were found between groups: the titanium group and the zirconia group did not differ significantly, but both demonstrated significantly smaller mean rotational freedoms than the alumina group (P < .05). Rotational freedom was less than 3 degrees for all abutments.
CONCLUSIONS: The hexagonal misfit of the Procera abutment on the implant hexagon may be implicated in screw joint loosening. In the present study, all types of CAD/CAM Procera abutments consistently showed less than 3 degrees of rotational freedom in a situation where the abutment was connected to an implant by a hexagonal external connection.

Entities:  

Mesh:

Substances:

Year:  2006        PMID: 16955608

Source DB:  PubMed          Journal:  Int J Oral Maxillofac Implants        ISSN: 0882-2786            Impact factor:   2.804


  9 in total

1.  Analytical and experimental position stability of the abutment in different dental implant systems with a conical implant-abutment connection.

Authors:  Wiebke Semper-Hogg; Silvan Kraft; Sebastian Stiller; Juergen Mehrhof; Katja Nelson
Journal:  Clin Oral Investig       Date:  2012-07-22       Impact factor: 3.573

2.  Comparison of fit accuracy and torque maintenance of zirconia and titanium abutments for internal tri-channel and external-hex implant connections.

Authors:  Hakimeh Siadat; Elaheh Beyabanaki; Niloufar Mousavi; Marzieh Alikhasi
Journal:  J Adv Prosthodont       Date:  2017-08-16       Impact factor: 1.904

3.  Reverse torque evaluation in indexed and nonindexed abutments of Morse Taper implants in a mechanical fatigue test.

Authors:  Caio Marques Martins; Elimário Venturin Ramos; Simone Kreve; Geraldo Alberto Pinheiro de Carvalho; Aline Batista Gonçalves Franco; Luís Guilherme Scavone de Macedo; Alecsandro de Moura Silva; Sérgio Candido Dias
Journal:  Dent Res J (Isfahan)       Date:  2019 Mar-Apr

4.  Rotational tolerances of a titanium abutment in the as-received condition and after screw tightening in a conical implant connection.

Authors:  Rosario Prisco; Giuseppe Troiano; Luigi Laino; Khrystyna Zhurakivska
Journal:  J Adv Prosthodont       Date:  2021-12-22       Impact factor: 1.904

5.  Effect of Different Surface Treatments on the Micro-Shear Bond Strength and Surface Characteristics of Zirconia: An In Vitro Study.

Authors:  Ann Sales; Shobha J Rodrigues; M Mahesh; Kishore Ginjupalli; Thilak Shetty; Umesh Y Pai; Sharon Saldanha; Puneeth Hegde; Sandipan Mukherjee; Vignesh Kamath; Prashant Bajantri; N Srikant; Ravindra Kotian
Journal:  Int J Dent       Date:  2022-04-14

6.  Rationale for the use of CAD/CAM technology in implant prosthodontics.

Authors:  Jaafar Abduo; Karl Lyons
Journal:  Int J Dent       Date:  2013-04-16

7.  Clinical Outcome of Inter-Proximal Papilla between a Tooth and a Single Implant Treated with CAD/CAM Abutments: a Cross-Sectional Study.

Authors:  Tiago Borges; Tiago Lima; Agata Carvalho; Vasco Carvalho
Journal:  J Oral Maxillofac Res       Date:  2012-10-01

Review 8.  Trends in computer-aided manufacturing in prosthodontics: a review of the available streams.

Authors:  Jaafar Abduo; Karl Lyons; Mohammed Bennamoun
Journal:  Int J Dent       Date:  2014-04-08

9.  Evaluation of removal forces of implant-supported zirconia copings depending on abutment geometry, luting agent and cleaning method during re-cementation.

Authors:  Matthias Rödiger; Sven Rinke; Fenja Ehret-Kleinau; Franziska Pohlmeyer; Katharina Lange; Ralf Bürgers; Nikolaus Gersdorff
Journal:  J Adv Prosthodont       Date:  2014-06-24       Impact factor: 1.904