Complications associated with an implantable vascular access device.

Patrick A Dillon1, Robert P Foglia.   

Abstract

PURPOSE: Implantable vascular access devices (ports) are well accepted in the management of many pediatric conditions. Modifications have improved port function, patient satisfaction, and enhanced compatibility with imaging studies. We reviewed our experience with a port system and identified unique mechanical complications.
METHODS: From 1998 to the present, 301 patients underwent 296 port insertions and 175 port removals. We assessed medical records, radiographs, and operative findings. The 6.6F MRI Low-Profile Implanted Port (Bard Access Systems, Salt Lake City, Utah) was used almost exclusively and was assembled by the operating surgeon. Outcome measures included port reservoir leakage, catheter dislodgment, and number of device days until complication. Ports were implanted for multiple medical problems including 74.2% in hematology/oncology patients.
RESULTS: For 296 port insertions, 15 complications (5.1%) were identified in 13 patients (mean age, 8.4 years). Eleven leaks (3.7%) in 9 patients were found, with 9 leaks resulting from needle perforation of the port base and 2 leaks seen at the catheter connection site. Average port duration was 425 days (range, 12-1266 days) before leakage. Four patients had catheter dislodgment (1.4%), with 3 of 4 catheters embolizing to the heart or pulmonary artery. Patients were asymptomatic, and catheters were retrieved by interventional radiology. Dislodgment at the catheter-port connection site was seen in 3 of 4 cases, and average port duration was 1075 days (range, 269-2657 days) until catheter separation. Twelve of 13 patients had successful implantation of a new port system.
CONCLUSIONS: This study identifies that (1) mechanical port complications (5.1%) are not rare for this device; (2) regardless of port age, the thin plastic base may result in a risk of perforation not seen in other devices; (3) the extended period before embolization likely indicates device wear rather than faulty assembly; and (4) complications could be successfully managed including retrieval of embolized catheters.

Entities:  

Mesh:

Year:  2006        PMID: 16952595     DOI: 10.1016/j.jpedsurg.2006.05.022

Source DB:  PubMed          Journal:  J Pediatr Surg        ISSN: 0022-3468            Impact factor:   2.545


  15 in total

1.  Tunable Nitric Oxide Release from S-Nitroso-N-acetylpenicillamine via Catalytic Copper Nanoparticles for Biomedical Applications.

Authors:  Jitendra Pant; Marcus J Goudie; Sean P Hopkins; Elizabeth J Brisbois; Hitesh Handa
Journal:  ACS Appl Mater Interfaces       Date:  2017-04-26       Impact factor: 9.229

2.  Incidence and outcome of retained Port-A-Cath fragments during removal.

Authors:  Olugbenga Michael Aworanti; Niall Linnane; Farhan Tareen; Alan Mortell
Journal:  Pediatr Surg Int       Date:  2017-06-05       Impact factor: 1.827

Review 3.  Totally implantable catheter migration and its percutaneous retrieval: case report and review of the literature.

Authors:  E Intagliata; F Basile; R Vecchio
Journal:  G Chir       Date:  2017 Sep-Oct

4.  Reduction in Thrombosis and Bacterial Adhesion with 7 Day Implantation of S-Nitroso-N-acetylpenicillamine (SNAP)-Doped Elast-eon E2As Catheters in Sheep.

Authors:  Elizabeth J Brisbois; Ryan P Davis; Anna M Jones; Terry C Major; Robert H Bartlett; Mark E Meyerhoff; Hitesh Handa
Journal:  J Mater Chem B       Date:  2015-02-28       Impact factor: 6.331

5.  Thromboresistance characterization of extruded nitric oxide-releasing silicone catheters.

Authors:  Kagya A Amoako; Christopher Archangeli; Hitesh Handa; Terry Major; Mark E Meyerhoff; Gail M Annich; Robert H Bartlett
Journal:  ASAIO J       Date:  2012 May-Jun       Impact factor: 2.872

6.  Analysis of risk factors for central venous port failure in cancer patients.

Authors:  Ching-Chuan Hsieh; Hsu-Huei Weng; Wen-Shih Huang; Wen-Ke Wang; Chiung-Lun Kao; Ming-Shian Lu; Chia-Siu Wang
Journal:  World J Gastroenterol       Date:  2009-10-07       Impact factor: 5.742

7.  Complicated vascular access port removals: incidence, antecedents and avoidance.

Authors:  Warwick J Teague; Dina Fouad; Fraser D Munro; Amanda J McCabe
Journal:  Pediatr Surg Int       Date:  2015-07-30       Impact factor: 1.827

8.  Attenuation of thrombosis and bacterial infection using dual function nitric oxide releasing central venous catheters in a 9day rabbit model.

Authors:  Elizabeth J Brisbois; Terry C Major; Marcus J Goudie; Mark E Meyerhoff; Robert H Bartlett; Hitesh Handa
Journal:  Acta Biomater       Date:  2016-08-06       Impact factor: 8.947

9.  [Catheters with implantable chamber: report of a series of 970 cases].

Authors:  El Hassane Kabiri; Massine El Hammoumi; Akram Traibi; Faycal El Oueriachi; Adil Arsalane
Journal:  Pan Afr Med J       Date:  2012-07-13

10.  Dislodgement of port-A catheters in pediatric oncology patients: 11 years of experience.

Authors:  Su-Chen Wang; Chia-Hui Tsai; Chiu-Ping Hou; Shin-Yi Lee; Sheung-Fat Ko; Chih-Chen Hsiao; Yu-Chieh Chen; Jiin-Haur Chuang; Jiunn-Ming Sheen
Journal:  World J Surg Oncol       Date:  2013-08-13       Impact factor: 2.754