| Literature DB >> 16940375 |
Abstract
A mathematical model has been developed for the assessment of patient skin doses from cardiac catheterization procedures. This uses exposure and projection data stored in the DICOM image files. Since these contain only information about the acquisition runs, a correction is needed to estimate and include the contribution from fluoroscopy. Maximum skin doses calculated by the model were found to correlate well with those measured on Kodak EDR2 film. Three methods for including the contribution from fluoroscopy were investigated, and all successfully identified patients receiving skin doses in excess of 1 Gy. It is hoped to automate this tool for routine assessment of skin doses in our cardiac catheterization laboratories.Entities:
Mesh:
Year: 2006 PMID: 16940375 DOI: 10.1259/bjr/57359387
Source DB: PubMed Journal: Br J Radiol ISSN: 0007-1285 Impact factor: 3.039