Evaluation of non-lead-based protective radiological material in spinal surgery.

Gaetano J Scuderi1, Georgiy V Brusovanik, David R Campbell, Robert P Henry, Brain Kwon, Alexander R Vaccaro.   

Abstract

BACKGROUND CONTEXT: Traditionally, lead-based garments are the standard method of intraoperative radiation protection during fluoroscopy. Unfortunately, the lead used is heavy, lacks durability, is difficult to launder, and its disposal is associated with environmental hazards.
PURPOSE: An evaluation of the protective radiation efficiency of three commercially available radiation protective garments compared with a standardized lead protective shield. STUDY DESIGN/
SETTING: Measured radiation transmission through lead and three commercially available lightweight radiological protective garments (Xenolite, EarthSafe, and Demron) was performed using a standard, calibrated dosimeter.
METHODS: Radiation transmission, attenuation, lead equivalencies as well as garment weight comparisons were measured. The tests were repeated through a range of voltage and tube current settings that are common to clinical radiological applications (60-120 keV).
RESULTS: All materials tested demonstrated effectiveness at common clinically relevant energy exposures (100 keV). EarthSafe and Xenolite demonstrated 0.5 mm lead equivalency protection at 80 and 100 keV X-ray energies but not at higher energy levels (>100 keV), which is where most radiological procedures are performed utilizing more advanced technological imaging equipment. Demron was best able to effectively shield ionizing radiation at higher energy levels (>100 keV). The lightweight nature of these lead-free materials may result in less fatigue and musculoskeletal complaints by the wearer.
CONCLUSIONS: Of the tested lead-free garments, Demron appears to offer equivalent levels of protection to standard lead-based shields within traditional energy zones but with less weight than standard lead-based shields.

Entities:  

Mesh:

Substances:

Year:  2006        PMID: 16934731     DOI: 10.1016/j.spinee.2005.09.010

Source DB:  PubMed          Journal:  Spine J        ISSN: 1529-9430            Impact factor:   4.166


  3 in total

1.  Protection evaluation of non-lead radiation-shielding fabric: preliminary exposure-dose study.

Authors:  Ju Hee Kang; Song Hee Oh; Jung-Il Oh; Seong-Hun Kim; Yong-Suk Choi; Eui-Hwan Hwang
Journal:  Oral Radiol       Date:  2018-06-25       Impact factor: 1.852

Review 2.  Strategies towards injectable, load-bearing materials for the intervertebral disc: a review and outlook.

Authors:  Cecilia Persson; Svante Berg
Journal:  J Mater Sci Mater Med       Date:  2012-09-29       Impact factor: 3.896

3.  Evaluation of scattered radiation dose received by medical staff during uterine artery embolization in the operating room.

Authors:  Hsien-Wen Chiang; Hsien-Jen Chiang; Jung-Hui Li; Leo Leung-Chit Tsang
Journal:  Technol Health Care       Date:  2020       Impact factor: 1.285

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.