Literature DB >> 16932037

Validation of three oscillometric blood pressure devices against auscultatory mercury sphygmomanometer in children.

Sik-Nin Wong1, Rita Yn Tz Sung, Lettie Chuk-Kwan Leung.   

Abstract

AIM: To validate Welch-Allyn Vital Sign Monitor, Dinamap Procare-120 and Datascope Accutorr Plus against auscultatory mercury sphygmomanometer in children aged 5-15 years old according to the International Protocol of European Society of Hypertension adapted for validation in children.
METHOD: One hundred and thirty two children were studied (44 for each device; 67 boys, 65 girls). Each underwent seven sequential BP measurements on the right arm resting in the sitting position, alternately with the mercury sphygmomanometer read simultaneously by two independent, trained observers and the test device by a third observer.
RESULTS: For the Welch-Allyn monitor, the mean+/-SD of differences (device minus auscultatory BP) were -4.39+/-4.82 mmHg for systolic blood pressure and -4.1+/-7.07 mmHg for diastolic blood pressure. The device failed phase 2.1 for both systolic blood pressure (55, 91 and 98% were within 5, 10 and 15 mmHg, respectively) and diastolic blood pressure measurements (46, 82 and 95% were within 5, 10 and 15 mmHg, respectively). For the Dinamap device, the mean+/-SD of differences were -3.08+/-5.21 mmHg for systolic blood pressure and -4.61+/-9.35 mmHg for diastolic blood pressure. The device passed phase 2.1 for systolic blood pressure (71, 96 and 98% were within 5, 10 and 15 mmHg, respectively) but failed for diastolic blood pressure (51, 72 and 91% were within 5, 10, and 15 mmHg, respectively). For the Datascope device, mean+/-SD of differences were -0.9+/-4.33 mmHg for systolic blood pressure and -1.20+/-6.48 mmHg for diastolic blood pressure. The device passed phase 2.1 in that 84, 97 and 99% of systolic blood pressure, and 61, 89 and 97% of diastolic blood pressure readings were within 5, 10 and 15 mmHg, respectively. It also passed phase 2.2 for both systolic blood pressure and diastolic blood pressure.
CONCLUSION: We performed an independent validation of three oscillometric BP devices in children. Overall Datascope Accutorr Plus passed, whereas Welch-Allyn Vital Sign Monitor and Dinamap Procare-120 failed an adapted IP-ESH.

Entities:  

Mesh:

Substances:

Year:  2006        PMID: 16932037     DOI: 10.1097/01.mbp.0000209082.09623.b4

Source DB:  PubMed          Journal:  Blood Press Monit        ISSN: 1359-5237            Impact factor:   1.444


  51 in total

1.  High prevalence of elevated blood pressure among children with neurofibromatosis type 1.

Authors:  Tom Dubov; Hagit Toledano-Alhadef; Gil Chernin; Shlomi Constantini; Roxana Cleper; Shay Ben-Shachar
Journal:  Pediatr Nephrol       Date:  2015-08-28       Impact factor: 3.714

2.  Evaluation of the Omron MX3 Plus monitor for blood pressure measurement in adolescents.

Authors:  Diego Giulliano Destro Christofaro; Juliano Casonatto; Marcos Doederlein Polito; Jefferson Rosa Cardoso; Rômulo Fernandes; Débora Alves Guariglia; Aline Mendes Gerage; Arli Ramos de Oliveira
Journal:  Eur J Pediatr       Date:  2009-02-17       Impact factor: 3.183

3.  Influence of breastfeeding on retinal vessel calibers in school-age children. The Generation R Study.

Authors:  O Gishti; V W V Jaddoe; L Duijts; O H Franco; A Hofman; M K Ikram; R Gaillard
Journal:  Eur J Clin Nutr       Date:  2015-07-15       Impact factor: 4.016

4.  Associations of maternal and fetal vitamin D status with childhood body composition and cardiovascular risk factors.

Authors:  Kozeta Miliku; Janine F Felix; Trudy Voortman; Henning Tiemeier; Darryl W Eyles; Thomas H Burne; John J McGrath; Vincent W V Jaddoe
Journal:  Matern Child Nutr       Date:  2018-09-21       Impact factor: 3.092

5.  Validation of Datascope Accutorr Plus™ using British Hypertension Society (BHS) and Association for the Advancement of Medical Instrumentation (AAMI) protocol guidelines.

Authors:  Rajab Ali Khawaja; Riaz Qureshi; Abdul Hay Mansure; Mohammad Emran Yahya
Journal:  J Saudi Heart Assoc       Date:  2010-03-10

6.  Maternal prepregnancy BMI or weight and offspring's blood pressure: Systematic review.

Authors:  Helena Ludwig-Walz; Milan Schmidt; Anke L B Günther; Anja Kroke
Journal:  Matern Child Nutr       Date:  2017-11-23       Impact factor: 3.092

7.  Childhood kidney outcomes in relation to fetal blood flow and kidney size.

Authors:  Marjolein N Kooijman; Hanneke Bakker; Albert J van der Heijden; Albert Hofman; Oscar H Franco; Eric A P Steegers; H Rob Taal; Vincent W V Jaddoe
Journal:  J Am Soc Nephrol       Date:  2014-05-08       Impact factor: 10.121

8.  Attention deficit-hyperactivity disorder is associated with reduced blood pressure and serum vitamin D levels: results from the nationwide German Health Interview and Examination Survey for Children and Adolescents (KiGGS).

Authors:  Thomas Meyer; Andreas Becker; Jessika Sundermann; Aribert Rothenberger; Christoph Herrmann-Lingen
Journal:  Eur Child Adolesc Psychiatry       Date:  2016-05-19       Impact factor: 4.785

9.  Comparison of mercury sphygmomanometry blood pressure readings with oscillometric and central blood pressure in predicting target organ damage in youth.

Authors:  Elaine M Urbina; Philip R Khoury; Connie E McCoy; Stephen R Daniels; Lawrence M Dolan; Thomas R Kimball
Journal:  Blood Press Monit       Date:  2015-06       Impact factor: 1.444

10.  Evaluation of risk of falls and orthostatic hypotension in older, long-term topical beta-blocker users.

Authors:  Wishal D Ramdas; Nathalie van der Velde; Tischa J M van der Cammen; Roger C W Wolfs
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2009-05-19       Impact factor: 3.117

View more