Literature DB >> 16931676

Change in expiratory flow detects partial endotracheal tube obstruction in pressure-controlled ventilation.

Rafael Kawati1, Laszlo Vimlati, Josef Guttmann, Göran Hedenstierna, Ulf Sjöstrand, Stefan Schumann, Michael Lichtwarck-Aschoff.   

Abstract

Only extreme degrees of endotracheal tube (ETT) narrowing can be detected with monitoring of tidal volume (V(T)) during pressure-controlled ventilation (PCV). To assess the degree of ETT obstruction in PCV and to compare it to V(T) monitoring, we produced 3 levels of partial ETT obstruction in 11 healthy anesthetized piglets using ETTs of 4 different inner diameters (IDs 9.0, 8.0, 7.0, and 6.0 mm). An expiratory flow over volume ((e)-V) curve was plotted and the time constant (tau(e)) at 15% of expiration time (T(e)) was calculated. We also calculated the fractional volume expired during the first 15% of T(e) (V(ex fract,15)) and compared those variables to full expiratory V(T) for each of the 3 obstructions. V(T) monitoring failed to detect ETT narrowing. By contrast, V(ex fract,15) decreased and tau(e) increased significantly with increasing ETT narrowing (for IDs 9.0, 8.0, 7.0, and 6.0, mean V(ex fract,15) was 195, 180, 146, and 134 mL respectively and mean tau(e) was 380, 491, 635, 794 ms for IDs 9.0, 8.0, 7.0, and 6.0 respectively). We conclude that when the elastic recoil that drives (e) is appropriately considered, analysis of (e) and V(ex fract,15) detects partial ETT obstruction during PCV.

Entities:  

Mesh:

Year:  2006        PMID: 16931676     DOI: 10.1213/01.ane.0000229777.80271.4a

Source DB:  PubMed          Journal:  Anesth Analg        ISSN: 0003-2999            Impact factor:   5.108


  2 in total

1.  Sound level analysis in endotracheal tube obstruction in spontaneous breathing and mechanical ventilation-an animal model study.

Authors:  Behzad Maghsoodi; Golnar Sabetian; Aram Azimi; Nader Tanideh; Alireza Mehdizade
Journal:  J Clin Monit Comput       Date:  2016-12-26       Impact factor: 2.502

2.  Sound analysis in an in vitro endotracheal tube model.

Authors:  Young Sik Park; Young Wook Kee; Kwang Suk Park; Jinwoo Lee; Sang-Min Lee; Jae-Joon Yim; Chul-Gyu Yoo; Young Whan Kim; Sung Koo Han; Seok-Chul Yang
Journal:  Korean J Intern Med       Date:  2011-11-28       Impact factor: 2.884

  2 in total