Effects of central corneal thickness and corneal curvature on the intraocular pressure measurement by Goldmann applanation tonometer and ocular blood flow pneumatonometer.

Tarek A Saleh1, Morag Adams, Bill McDermott, Kate G Claridge, Paul Ewings.   

Abstract

PURPOSE: To assess the effects of central corneal thickness (CCT) and corneal curvature (CC) on the measurements of intraocular pressure (IOP) using Goldmann applanation tonometer (GAT) and the ocular blood flow pneumatonometer (OBFT).
METHODS: 104 patients were recruited from a glaucoma clinic. The CCT was measured using ultrasound pachymetry and the mean radius of CC using a keratometer. The IOP of each eye was measured using both GAT and the OBFT in a random order. Right eyes only were analysed for statistical purposes.
RESULTS: The mean (+/-SD) IOP by GAT and OBFT was 18.2 mmHg (+/-4.4) and 18.2 mmHg (+/-4.0), respectively, with no statistically significant difference. IOP measurement with both instruments varied with CCT and CC. GAT showed an IOP increase of 0.40 mmHg per 10 microm increase of CCT and OBFT showed an increase of 0.38 mmHg in IOP per 10 microm increase of CCT. Multiple regression analysis showed that the effect of CCT was statistically significant (P<0.001) on IOP recorded by both the GAT and OBFT but CC did not have a statistically significant effect on IOP recordings performed by either technique.
CONCLUSION: IOP measurements by GAT and OBFT are positively correlated with CCT with both tonometers being similarly affected. There was no significant correlation between CC and IOP measured by either tonometer.

Entities:  

Mesh:

Year:  2006        PMID: 16925697     DOI: 10.1111/j.1442-9071.2006.01266.x

Source DB:  PubMed          Journal:  Clin Exp Ophthalmol        ISSN: 1442-6404            Impact factor:   4.207


  8 in total

1.  Intraocular pressure and ocular pulse amplitude variations during the Valsalva maneuver.

Authors:  Umit Aykan; Mesut Erdurmus; Burhan Yilmaz; Ahmet Hamdi Bilge
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2010-03-24       Impact factor: 3.117

2.  Effects of phacoemulsification surgery on ocular hemodynamics.

Authors:  Adem Turk; Suleyman Mollamehmetoglu; Halil Ibrahim Imamoglu; Mehmet Kola; Hidayet Erdol; Nurettin Akyol
Journal:  Int J Ophthalmol       Date:  2013-08-18       Impact factor: 1.779

3.  Effects of corneal thickness on intraocular pressure measured with three different tonometers.

Authors:  Hiroki Murase; Akira Sawada; Kiyofumi Mochizuki; Tetsuya Yamamoto
Journal:  Jpn J Ophthalmol       Date:  2009-01-30       Impact factor: 2.447

4.  Effects of central corneal thickness, central corneal power, and axial length on intraocular pressure measurement assessed with goldmann applanation tonometry.

Authors:  Fatih Özcura; Sayime Aydın; Nevin Uzgören
Journal:  Jpn J Ophthalmol       Date:  2008-11-11       Impact factor: 2.447

5.  Comparison of iCare tonometer and Goldmann applanation tonometry in normal corneas and in eyes with automated lamellar and penetrating keratoplasty.

Authors:  M L Salvetat; M Zeppieri; F Miani; C Tosoni; L Parisi; P Brusini
Journal:  Eye (Lond)       Date:  2011-03-25       Impact factor: 3.775

6.  Effect of corneal astigmatism on intraocular pressure measurement using ocular response analyzer and Goldmann applanation tonometer.

Authors:  Mana Hagishima; Kazutaka Kamiya; Fusako Fujimura; Tetsuya Morita; Nobuyuki Shoji; Kimiya Shimizu
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2009-09-29       Impact factor: 3.117

7.  Purinergic dysregulation causes hypertensive glaucoma-like optic neuropathy.

Authors:  Youichi Shinozaki; Kenji Kashiwagi; Kazuhiko Namekata; Akiko Takeda; Nobuhiko Ohno; Bernard Robaye; Takayuki Harada; Takeshi Iwata; Schuichi Koizumi
Journal:  JCI Insight       Date:  2017-10-05

8.  Detecting IOP Fluctuations in Glaucoma Patients.

Authors:  Brenda Nuyen; Kaweh Mansouri
Journal:  Open Ophthalmol J       Date:  2016-02-29
  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.