An algorithm for continuous real-time QT interval monitoring.

Eric D Helfenbein1, Sophia H Zhou, James M Lindauer, Dirk Q Field, Richard E Gregg, John J Wang, Scott S Kresge, Francis P Michaud.   

Abstract

QT interval measurement in the patient monitoring environment is receiving much interest because of the potential for proarrhythmic effects from both cardiac and noncardiac drugs. The American Heart Association and American Association of Critical Care Nurses practice standards for ECG monitoring in hospital settings now recommend frequent monitoring of QT interval when patients are started on a potentially proarrhythmic drug. We developed an algorithm to continuously measure QT interval in real-time in the patient monitoring setting. This study reports our experience in developing and testing this automated QT algorithm. Compared with the environment of resting ECG analysis, real-time ECG monitoring has a number of challenges: significantly more amounts of muscle and motion artifact, increased baseline wander, a varied number and location of ECG leads, and the need for trending and for alarm generation when QT interval prolongation is detected. We have used several techniques to address these challenges. In contiguous 15-second time windows, we average the signal of tightly clustered normal beats detected by a real-time arrhythmia-monitoring algorithm to minimize the impact of artifact. Baseline wander is reduced by zero-phase high-pass filtering and subtraction of isoelectric points as determined by median signal values in a localized region. We compute a root-mean-squared ECG waveform from all available leads and use a novel technique to measure the QT interval. We have tested this algorithm against standard and proprietary ECG databases. Our real-time QT interval measurement algorithm proved to be stable, accurate, and able to track changing QT values.

Entities:  

Mesh:

Year:  2006        PMID: 16920145     DOI: 10.1016/j.jelectrocard.2006.05.018

Source DB:  PubMed          Journal:  J Electrocardiol        ISSN: 0022-0736            Impact factor:   1.438


  9 in total

1.  Philips QT interval measurement algorithms for diagnostic, ambulatory, and patient monitoring ECG applications.

Authors:  Sophia H Zhou; Eric D Helfenbein; James M Lindauer; Richard E Gregg; Dirk Q Feild
Journal:  Ann Noninvasive Electrocardiol       Date:  2009-01       Impact factor: 1.468

Review 2.  Diagnostic performance of electronic syndromic surveillance systems in acute care: a systematic review.

Authors:  M Kashiouris; J C O'Horo; B W Pickering; V Herasevich
Journal:  Appl Clin Inform       Date:  2013-05-08       Impact factor: 2.342

3.  Grasping trajectories in a virtual environment adhere to Weber's law.

Authors:  Aviad Ozana; Sigal Berman; Tzvi Ganel
Journal:  Exp Brain Res       Date:  2018-04-16       Impact factor: 1.972

4.  Prevention of torsade de pointes in hospital settings: a scientific statement from the American Heart Association and the American College of Cardiology Foundation.

Authors:  Barbara J Drew; Michael J Ackerman; Marjorie Funk; W Brian Gibler; Paul Kligfield; Venu Menon; George J Philippides; Dan M Roden; Wojciech Zareba
Journal:  Circulation       Date:  2010-02-08       Impact factor: 29.690

Review 5.  Prevention of torsade de pointes in hospital settings: a scientific statement from the American Heart Association and the American College of Cardiology Foundation.

Authors:  Barbara J Drew; Michael J Ackerman; Marjorie Funk; W Brian Gibler; Paul Kligfield; Venu Menon; George J Philippides; Dan M Roden; Wojciech Zareba
Journal:  J Am Coll Cardiol       Date:  2010-03-02       Impact factor: 24.094

Review 6.  Effect of Hydroxychloroquine on QTc in Patients Diagnosed with COVID-19: A Systematic Review and Meta-Analysis.

Authors:  Angelos Arfaras-Melainis; Andreas Tzoumas; Damianos G Kokkinidis; Maria Salgado Guerrero; Dimitrios Varrias; Xiaobo Xu; Luis Cerna; Ricardo Avendano; Cameron Kemal; Leonidas Palaiodimos; Robert T Faillace
Journal:  J Cardiovasc Dev Dis       Date:  2021-05-13

7.  New approach for T-wave end detection on electrocardiogram: performance in noisy conditions.

Authors:  Carlos R Vázquez-Seisdedos; João Evangelista Neto; Enrique J Marañón Reyes; Aldebaro Klautau; Roberto C Limão de Oliveira
Journal:  Biomed Eng Online       Date:  2011-09-09       Impact factor: 2.819

8.  QT interval prolongation and torsade de pointes in patients with COVID-19 treated with hydroxychloroquine/azithromycin.

Authors:  Ehud Chorin; Lalit Wadhwani; Silvia Magnani; Matthew Dai; Eric Shulman; Charles Nadeau-Routhier; Robert Knotts; Roi Bar-Cohen; Edward Kogan; Chirag Barbhaiya; Anthony Aizer; Douglas Holmes; Scott Bernstein; Michael Spinelli; David S Park; Carugo Stefano; Larry A Chinitz; Lior Jankelson
Journal:  Heart Rhythm       Date:  2020-05-12       Impact factor: 6.343

9.  Why we should be more careful using hydroxychloroquine in influenza season during COVID-19 pandemic?

Authors:  Hülya Gamze Çelik; Şiran Keske; Ülker Şener; Müge Tekbaş; Mahir Kapmaz; Şükrü Taylan Şahin; Aslı Özyıldırım; Saide Aytekin; Vedat Aytekin; Önder Ergönül
Journal:  Int J Infect Dis       Date:  2020-10-29       Impact factor: 3.623

  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.