| Literature DB >> 16891524 |
Bram M W Diederen1, Marcel F Peeters.
Abstract
We evaluated two new immunochromatographic assays for their abilities to detect Legionella pneumophila serogroup 1 antigen in urine. The results were compared with those obtained by the Binax NOW urinary antigen test. The sensitivities and specificities were estimated to be 71.2% and 96.6%, respectively, for the Rapid U test; 31.5% and 98.9%, respectively, for the SD Bioline test; and 91.8% and 100%, respectively, for the Binax NOW test.Entities:
Mesh:
Substances:
Year: 2006 PMID: 16891524 PMCID: PMC1594613 DOI: 10.1128/JCM.00799-06
Source DB: PubMed Journal: J Clin Microbiol ISSN: 0095-1137 Impact factor: 5.948