Literature DB >> 16872969

Structural deterioration of the Freestyle aortic valve: mode of presentation and mechanisms.

Siamak Mohammadi1, Richard Baillot, Pierre Voisine, Patrick Mathieu, François Dagenais.   

Abstract

BACKGROUND: Structural valve deterioration is the major cause of bioprosthetic valve failure. Because of the unique design features and anti-calcification treatment of the Freestyle (Medtronic Inc, Minneapolis, Minn) stentless bioprosthesis, development of structural valve deterioration may differ in comparison with other bioprosthetic valves. This study evaluates the mechanisms and clinical presentation of structural valve deterioration in the Freestyle stentless bioprosthesis.
METHODS: Between January 1993 and August 2005, 608 patients underwent aortic valve replacement with a Freestyle stentless bioprosthesis. The implantation technique was subcoronary in 475 patients and a root replacement in 133 patients. Mean overall follow-up was 5.6 +/- 3.4 years. Follow-up was complete in all patients. Clinical and echocardiographic follow-ups were conducted prospectively.
RESULTS: Freedom from structural valve deterioration was 95.8% at 10 years. Twelve patients showed evidence of structural valve deterioration and underwent reoperation for aortic regurgitation (n = 10) or aortic stenosis (n = 2). The mean age of patients with structural valve deterioration was significantly lower than patients without structural valve deterioration (62.6 +/- 8.2 years vs 68.6 +/- 8.3 years, P = .02). The median time between implantation and explantation was 8.7 years (range: 1.9-13.3 years). Eleven structural valve deteriorations occurred after subcoronary implantation, and 1 structural valve deterioration occurred after root implantation (P = .4). The mechanisms of structural valve deterioration were leaflet tears in 10 patients (6 in the left coronary cusp and 4 in the right coronary cusp), severe valve calcification in 1 patient, and cusp fibrosis in 1 patient. The interval between onset of symptoms and reoperation was acute or subacute in 10 patients.
CONCLUSION: At 10 years, the Freestyle stentless bioprosthesis shows excellent freedom from structural valve deterioration. Structural valve deterioration in the Freestyle stentless bioprosthesis relates to leaflet tear with minimal calcification in the majority of cases. Because of the fast onset of symptoms with leaflet tear, patients with a Freestyle stentless bioprosthesis should be informed of the preferential mode of failure and time-frame of symptoms.

Entities:  

Mesh:

Year:  2006        PMID: 16872969     DOI: 10.1016/j.jtcvs.2006.03.056

Source DB:  PubMed          Journal:  J Thorac Cardiovasc Surg        ISSN: 0022-5223            Impact factor:   5.209


  12 in total

1.  Valvular disease: The private life of tissue valves.

Authors:  Magdi H Yacoub; Ismail El-Hamamsy
Journal:  Nat Rev Cardiol       Date:  2010-08       Impact factor: 32.419

2.  Torn cusp in a freestyle stentless bioprosthesis 5 years after implantation: report of a case.

Authors:  Shinya Fukui; Motonobu Nishimura; Goro Matsumiya; Masao Yoshitatsu; Hajime Matsue; Yoshiki Sawa
Journal:  Surg Today       Date:  2007-10-25       Impact factor: 2.549

3.  The Quebec Heart Institute: 50 years of excellence in cardiology.

Authors:  Gilles R Dagenais; François Philippon; Jean-Pierre Després; Jean G Dumesnil; Paul Cartier; Peter M Bogaty; Michel Lemieux; André Moisan
Journal:  Can J Cardiol       Date:  2007-10       Impact factor: 5.223

4.  Influence of Age on Longevity of a Stentless Aortic Valve.

Authors:  Bo Yang; Aroosa Malik; Linda Farhat; Alexander Makkinejad; Elizabeth L Norton; Mohamed-Ali Sareini; Emma St Pierre; Xiaoting Wu; Jonathan Haft; Matthew Romano; Steven F Bolling; Himanshu J Patel; G Michael Deeb
Journal:  Ann Thorac Surg       Date:  2019-12-23       Impact factor: 4.330

Review 5.  Management of acute regurgitation in left-sided cardiac valves.

Authors:  Nahush A Mokadam; Karen K Stout; Edward D Verrier
Journal:  Tex Heart Inst J       Date:  2011

6.  Pseudoaneurysm formation after medtronic freestyle porcine aortic bioprosthesis implantation: a word of caution.

Authors:  Brian R Englum; Elizabeth N Pavlisko; Molly C Mack; Asvin M Ganapathi; Matthew A Schechter; Jennifer M Hanna; G Chad Hughes
Journal:  Ann Thorac Surg       Date:  2014-10-07       Impact factor: 4.330

7.  Total ellipse of the heart valve: the impact of eccentric stent distortion on the regional dynamic deformation of pericardial tissue leaflets of a transcatheter aortic valve replacement.

Authors:  Paul S Gunning; Neelakantan Saikrishnan; Ajit P Yoganathan; Laoise M McNamara
Journal:  J R Soc Interface       Date:  2015-12-06       Impact factor: 4.118

Review 8.  Stentless aortic valve replacement: an update.

Authors:  Junjiro Kobayashi
Journal:  Vasc Health Risk Manag       Date:  2011-06-02

9.  Early Doppler-echocardiography evaluation of Carpentier-Edwards Standard and Carpentier-Edwards Magna aortic prosthetic valve: comparison of hemodynamic performance.

Authors:  Giovanni Minardi; Giovanni Pulignano; Donatella Del Sindaco; Martina Sordi; Herribert Pavaci; Amedeo Pergolini; Giordano Zampi; Francesca Moschella Orsini; Carlo Gaudio; Francesco Musumeci
Journal:  Cardiovasc Ultrasound       Date:  2011-11-24       Impact factor: 2.062

10.  Structural valve deterioration after aortic valve replacement with Medtronic freestyle stentless porcine aortic root bioprostheses.

Authors:  Hidefumi Nishida; Tatsuhiko Komiya; Genichi Sakaguchi; Takeshi Shimamoto
Journal:  J Card Surg       Date:  2013-10-22       Impact factor: 1.620

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.