Literature DB >> 16844406

Monitoring of scoliosis surgery with epidurally recorded motor evoked potentials (D wave) revealed false results.

S Ulkatan1, M Neuwirth, F Bitan, C Minardi, A Kokoszka, V Deletis.   

Abstract

OBJECTIVE: To elucidate the mechanism behind D wave amplitude changes after surgical correction of scoliosis.
METHODS: We collected D wave and muscle MEP data from 93 patients (78 female, 15 male, age range 4-19 years, mean age 15.9 years), who underwent surgical correction of scoliosis. D waves were recorded via a catheter electrode inserted epidurally through the flavectomy. Muscle MEPs from lower limb muscles were also recorded. Muscle MEPs/D wave were elicited by short trains/single transcranial electrical stimuli. SEPs were elicited through bilateral percutaneous stimulation of the tibial nerves at the ankle and an averaged response from 100 to 200 single sweeps were recorded over the scalp at Cz'/Fz. In addition, we analyzed intraoperatively obtained X-ray images of the spine in 9 patients and preoperative spinal MRI in two of those nine.
RESULTS: After surgical correction of scoliosis in 25 of 93 (27%) patients, the D wave amplitude changed by more than 20% of its baseline value. A decremental change occurred in 21 (84%) and an incremental change in 4 (16%) patients. D wave decrements of more than 50% were observed in 5 patients without significant SEP changes in any of these cases. In 9 patients, intraoperatively obtained X-rays of the spine (before and after correction of spine curvature) showed no catheter displacement. Muscle MEPs did not change and postoperative sensory-motor status was normal. In 2 patients, preoperative MRI revealed displacement of the spinal cord towards the concave side of the scoliotic curvature.
CONCLUSIONS: During scoliosis surgery, D wave amplitude changes should be interpreted cautiously until the definitive cause(s) of these changes are found. One possible mechanism to explain D wave changes during scoliosis correction could involve rotation of the spinal cord within the spinal canal, and the relative position of the epidural recording catheter (ERC). Rotation of the spinal cord after correction of scoliosis could introduce a new relationship between the ERC and the corticospinal tracts (CTs). Due to high incidence of false D wave amplitude changes we suggest that this methodology should not be used to assess the functional integrity of the CTs during scoliosis surgery. SIGNIFICANCE: This study provides new insight into the methodology of D wave monitoring as well as strong evidence of a high incidence of false positive results using D wave monitoring during surgical correction of scoliosis.

Entities:  

Mesh:

Year:  2006        PMID: 16844406     DOI: 10.1016/j.clinph.2006.05.021

Source DB:  PubMed          Journal:  Clin Neurophysiol        ISSN: 1388-2457            Impact factor:   3.708


  14 in total

Review 1.  Intraoperative neurophysiological monitoring in spinal surgery.

Authors:  Jong-Hwa Park; Seung-Jae Hyun
Journal:  World J Clin Cases       Date:  2015-09-16       Impact factor: 1.337

Review 2.  Intraoperative motor evoked potential monitoring: overview and update.

Authors:  David B Macdonald
Journal:  J Clin Monit Comput       Date:  2006-07-11       Impact factor: 2.502

3.  Basic methodological principles of multimodal intraoperative monitoring during spine surgeries.

Authors:  Vedran Deletis
Journal:  Eur Spine J       Date:  2007-07-11       Impact factor: 3.134

Review 4.  Current opinions and recommendations on multimodal intraoperative monitoring during spine surgeries.

Authors:  Martin Sutter; Vedran Deletis; Jiri Dvorak; Andreas Eggspuehler; Dieter Grob; David Macdonald; Alfred Mueller; Francesco Sala; Tetsuya Tamaki
Journal:  Eur Spine J       Date:  2007-08-15       Impact factor: 3.134

5.  Intraoperative neurophysiological monitoring during complex spinal deformity cases in pediatric patients: methodology, utility, prognostication, and outcome.

Authors:  James Drake; Reinhard Zeller; Abhaya V Kulkarni; Samuel Strantzas; Laura Holmes
Journal:  Childs Nerv Syst       Date:  2010-03-07       Impact factor: 1.475

6.  Usefulness of multi-channels in intraoperative spinal cord monitoring: multi-center study by the Monitoring Committee of the Japanese Society for Spine Surgery and Related Research.

Authors:  Zenya Ito; Yukihiro Matsuyama; Kenichi Shinomiya; Muneharu Ando; Shigenori Kawabata; Tsukasa Kanchiku; Takanori Saito; Masato Takahashi; Shinichiro Taniguchi; Naoya Yamamoto; Kei Yamada; Kazunobu Kida; Yasushi Fujiwara; Sho Kobayashi; Kazuhiko Satomi; Toshikazu Tani
Journal:  Eur Spine J       Date:  2013-04-04       Impact factor: 3.134

7.  Intraoperative neuromonitoring practice patterns in spinal deformity surgery: a global survey of the Scoliosis Research Society.

Authors:  Pinar Yalinay Dikmen; Matthew F Halsey; Altug Yucekul; Marinus de Kleuver; Lloyd Hey; Peter O Newton; Irem Havlucu; Tais Zulemyan; Caglar Yilgor; Ahmet Alanay
Journal:  Spine Deform       Date:  2020-11-23

Review 8.  Current approach on spinal cord monitoring: the point of view of the neurologist, the anesthesiologist and the spine surgeon.

Authors:  Thomas N Pajewski; Vincent Arlet; Lawrence H Phillips
Journal:  Eur Spine J       Date:  2007-07-10       Impact factor: 3.134

9.  Four-limb muscle motor evoked potential and optimized somatosensory evoked potential monitoring with decussation assessment: results in 206 thoracolumbar spine surgeries.

Authors:  David B Macdonald; Zayed Al Zayed; Abdulmoneam Al Saddigi
Journal:  Eur Spine J       Date:  2007-07-19       Impact factor: 3.134

Review 10.  Intraoperative Spinal Cord Monitoring: Focusing on the Basic Knowledge of Orthopedic Spine Surgeon and Neurosurgeon as Members of a Team Performing Spine Surgery under Neuromonitoring.

Authors:  Tetsuya Tamaki; Muneharu Ando; Yukihiro Nakagawa; Hiroshi Iwasaki; Shunji Tsutsui; Masanari Takami; Hiroshi Yamada
Journal:  Spine Surg Relat Res       Date:  2021-03-10
View more