| Literature DB >> 16839735 |
Frank Gessler1, Katrin Hampe, Michaela Schmidt, Helge Böhnel.
Abstract
An immunomagnetic beads assay for the simultaneous quantification of botulinum neurotoxin types C and D was developed. Specific monoclonal antibodies against the heavy chain of the toxin and affinity-purified biotinylated polyclonal antibodies (pAbs) were used. The antibodies were preincubated with the sample. The complex being formed was then captured by magnetic beads coated with antimouse IgG. Streptavidin-poly-horseradish peroxidase, a signal amplifier, bound to the biotinylated pAb. A maximum sensitivity of approximately 0.3 minimal lethal doses for mice per milliliter was achieved with culture supernatants of both toxin types.Entities:
Mesh:
Substances:
Year: 2006 PMID: 16839735 DOI: 10.1016/j.diagmicrobio.2006.04.014
Source DB: PubMed Journal: Diagn Microbiol Infect Dis ISSN: 0732-8893 Impact factor: 2.803