An ultrasound mini-balance for measurement of therapy level ultrasound.

Yvonne Sutton1, Karne McBride, Stephen Pye.   

Abstract

This paper describes a cost-effective method for measuring acoustic power using a radiation force balance. The device is based around a long established balance design with a gantry arrangement fitted with an absorbing target. The notion of this balance design is that it can easily be constructed from materials that would be readily available within a clinical or industrial environment. The mini-balance was calibrated using a transfer standard against an NPL Reference balance, so a comparison of the performance between the two systems could be assessed. The measurements were completed at 1 MHz and 3 MHz and over the acoustic power range of 1 W to 15 W. The results show the acoustic power measured on the mini-balance to be within 5% of the reference measurements made on the NPL Balance. A separate systematic uncertainty budget is also presented based on studies made on the balance and on similar systems. The overall expanded uncertainty was calculated to be within 14% at 1 W level, decreasing with increasing power level to 7.4% above 5 W.

Entities:  

Mesh:

Year:  2006        PMID: 16825738     DOI: 10.1088/0031-9155/51/14/008

Source DB:  PubMed          Journal:  Phys Med Biol        ISSN: 0031-9155            Impact factor:   3.609


  3 in total

1.  Intense focused ultrasound can reliably induce sensations in human test subjects in a manner correlated with the density of their mechanoreceptors.

Authors:  Trevor C Dickey; Rowen Tych; Michel Kliot; John D Loeser; Kristin Pederson; Pierre D Mourad
Journal:  Ultrasound Med Biol       Date:  2011-11-21       Impact factor: 2.998

2.  Intense focused ultrasound as a potential research tool for the quantification of diurnal inflammatory pain.

Authors:  Josephine D Garcia; Michael Gofeld; P Ray Illian; John D Loeser; Michel Kliot; Abbi M McClintic; Alice Ward; Anning Yao; Pierre D Mourad
Journal:  Ultrasonics       Date:  2012-04-18       Impact factor: 2.890

3.  Intense focused ultrasound preferentially stimulates subcutaneous and focal neuropathic tissue: preliminary results.

Authors:  Abbi M McClintic; Trevor C Dickey; Michael Gofeld; Michel Kliot; John D Loeser; Philippe Richebe; Pierre D Mourad
Journal:  Pain Med       Date:  2012-11-08       Impact factor: 3.750

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.