Capillary liquid chromatography-microchip atmospheric pressure chemical ionization-mass spectrometry.

Pekka Ostman1, Sirkku Jäntti, Kestas Grigoras, Ville Saarela, Raimo A Ketola, Sami Franssila, Tapio Kotiaho, Risto Kostiainen.   

Abstract

A miniaturized nebulizer chip for capillary liquid chromatography-atmospheric pressure chemical ionization-mass spectrometry (capillary LC-microchip APCI-MS) is presented. The APCI chip consists of two wafers, a silicon wafer and a Pyrex glass wafer. The silicon wafer has a DRIE etched through-wafer nebulizer gas inlet, an edge capillary insertion channel, a stopper, a vaporizer channel and a nozzle. The platinum heater electrode and pads for electrical connection were patterned on to the Pyrex glass wafer. The two wafers were joined by anodic bonding, creating a microchip version of an APCI-source. The sample inlet capillary from an LC column is directly connected to the vaporizer channel of the APCI chip. The etched nozzle in the microchip forms a narrow sample plume, which is ionized by an external corona needle, and the formed ions are analyzed by a mass spectrometer. The nebulizer chip enables for the first time the use of low flow rate separation techniques with APCI-MS. The performance of capillary LC-microchip APCI-MS was tested with selected neurosteroids. The capillary LC-microchip APCI-MS provides quantitative repeatability and good linearity. The limits of detection (LOD) with a signal-to-noise ratio (S/N) of 3 in MS/MS mode for the selected neurosteroids were 20-1000 fmol (10-500 nmol l(-1)). LODs (S/N = 3) with commercial macro APCI with the same compounds using the same MS were about 10 times higher. Fast heat transfer allows the use of the optimized temperature for each compound during an LC run. The microchip APCI-source provides a convenient and easy method to combine capillary LC to any API-MS equipped with an APCI source. The advantages and potentials of the microchip APCI also make it a very attractive interface in microfluidic APCI-MS.

Entities:  

Mesh:

Substances:

Year:  2006        PMID: 16804601     DOI: 10.1039/b601290f

Source DB:  PubMed          Journal:  Lab Chip        ISSN: 1473-0189            Impact factor:   6.799


  2 in total

1.  Analysis of selective androgen receptor modulators by gas chromatography-microchip atmospheric pressure photoionization-mass spectrometry.

Authors:  Laura Luosujärvi; Markus Haapala; Mario Thevis; Ville Saarela; Sami Franssila; Raimo A Ketola; Risto Kostiainen; Tapio Kotiaho
Journal:  J Am Soc Mass Spectrom       Date:  2009-11-06       Impact factor: 3.109

2.  On-chip mass spectrometric analysis in non-polar solvents by liquid beam infrared matrix-assisted laser dispersion/ionization.

Authors:  Raphael D Urban; Tillmann G Fischer; Ales Charvat; Konstantin Wink; Benjamin Krafft; Stefan Ohla; Kirsten Zeitler; Bernd Abel; Detlev Belder
Journal:  Anal Bioanal Chem       Date:  2021-01-21       Impact factor: 4.142

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.