Q Q Zhang1, F T Djuth, Q F Zhou, C H Hu, J H Cha, K K Shung.
Abstract
A modified sol-gel method is used to prepare PZT thick film on Pt-coated silicon substrate. A new method of vacuum filling sol-gel precursor solution is introduced to improve film quality. The effects of the filling on PZT thick film structure and ferroelectric properties are discussed. The fabrication of a high frequency transducer with the PZT film as the actuating layer is described. The performance of the transducer is measured and results show that the transducer backed by E-Solder without a matching layer has a center frequency of 103 MHz and a bandwidth of 70%. Beam profile measurements show that the transducer has an axial resolution of 9.2 microm and a lateral resolution of 33 microm.Entities:
Mesh:
Substances:
Year: 2006 PMID: 16793087 DOI: 10.1016/j.ultras.2006.05.130
Source DB: PubMed Journal: Ultrasonics ISSN: 0041-624X Impact factor: 2.890